3D DSA Workflow
3D DSA Workflow
3D DSA Workflow Interventional Radiography: Cardiac 1. Prepare the patient, stand, C-arm and table. Check the injector, if used. Select a 3D DSAacquisition program. 55DR 2. Position the region of interest in the isocenter: Follow the screen prompts to ISO center AP and then to center lateral You can start immediately if the LED on the Isocenter key is lit Follow the "isocenter assistant" if the LED on the Isocenter key is flashing 3. Move C-arm to start position: Press the stand joystick down and deflect it Perform the 3D DSA test run to make sure that no collision will occur. 4. Perform a brief fluoroscopy for automatic setting of acquisition parameters: Press fluoro pedal of footswitch. Ready for 3D is displayed. 5. Perform the 3D DSA mask run. Start acquisition: Press the exposure release button or pedal and keep it pressed Movement is performed automatically. Mask images are acquired. The return run takes place automatically. 6. Automatic or manual injection. Start injector if manual injection is used. TEH 7. Perform the 3D DSA fill run: Rotational run will start automatically, if Auto is programmed. In- jection will also be performed automatically if the injector is set to "Armed" <-k- Start rotational run in time: Press the blue multifunctional button of the handswitch, if Manual is programmed Fill images are acquired. SIEMENS siemens.com/healthineers Healthineers . Job Aid | 3D DSA Workflow Effective Date: 08/15/2016 | Quality HOOD05162002274306 8. Let go the exposure release when all rotational runs are complete (acqui- sition is stopped automatically). Transfer of image data to the syngo Workplace: DIA Automatically in primary reconstruction Manually for secondary reconstruction Siemens Healthineers Services Page 2 of 3 Job Aid | 3D DSA Workflow Effective Date: 08/15/2016 | Quality HOOD05162002274306 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright @ Siemens Healthcare GmbH Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 3 of 3
- Artis Q
- Artis zee biplane
- Artis zee ceiling mounted
- Artis zee floor mounted
- Artis zeego
- Artis Zen
- syngo X Workplace