Access-i Remote Control Job Aid
The Access-i Remote Control Job Aid reviews the workflow steps for a user to operate a MR MAGNETOM system from another computer in addition to the syngo Acquisition Workplace.
Access-i Remote Control Job Aid Access-i Remote Control is an optional Feature. It allows the user to operate a MAGNETOM system from another computer in addition to the syngo Acquisition Workplace. Please Note: This software feature is not available for MAGNETOM Skyra-fit and MAGNETOM Avanto-fit MR Systems. You can control and edit the MR System examination programs from a non-Siemens Healthineers device. These devices can include Tablets, Computers, or Smartphones, independent of their platform, programming language, and operating system. You can access the syngo Acquisition workplace from the client in two ways: Passive: Allows you to view and receive acquired MR images on the client. All interactions are performed on the syngo Acquisition workplace or by another active client. Active: When using Active, you can control the measurements from the client; however, interactions at the syngo acquisition workplace are not possible. On the client, you can only operate programs that are “enabled” for remote access. There are two steps that need to be performed: Enter [Remote] in the Description field of the Directory Properties of the program in question. Please Note: The square brackets must be included. Directory Properties X XA30 User Tree > Brain > clinical libraries >> general Name aeneral Description [Remote] Body Part Examined Manage Table Positioning Strategy Archive Visibility Show irrespective of licensing OK Cancel Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 05 MAR 2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 10 SEP 2020 | HOOD05162003113055 USA – Effective Date: 11.24.2020 | HOOD05162003144668 Healthineers Job Aid | Access-i Remote Control Effective Date: 10 SEP 2020 | HOOD05162003113055 USA - Effective Date: 11.24.2020 | HOOD05162003144668 Add the Access-I-Dot add-in to the relevant program step. A program step that is “enabled” for remote access is called a “Template”. When you add this program step to the measurement queue, the expression [Template] is added as a prefix to the name of the step. The behavior of a template step is also quite different. As soon as you start the template step, a copy of this step is created and started, so that the original template step is never completed. This allows you to remotely control and edit the template step several times. Dot Add-In Configurator ? X 7 6 Guidance Enable Interactivity Queue GSP Series Label AutoCoverage AutoTiming Close To enable a measurement program for remote access, the directory properties need to be adjusted. Open the Dot Cockpit and select the Explorer task, then the Browser subtask Next select Measurement Program (Directory), in the tree view Right-click and choose Edit Properties from the context menu The Directory Properties dialog window opens In the Description field of Directory Properties window, enter [Remote] Please Note: the square brackets must be included. Click OK to confirm the entry You can remotely add this program to the measurement queue. To control the program steps, you must additionally add the Access-i Dot add-in. Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 05 MAR 2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 10 SEP 2020 | HOOD05162003113055 USA – Effective Date: 11.24.2020 | HOOD05162003144668 Healthineers Page 2 of 5 Job Aid | Access-i Remote Control Effective Date: 10 SEP 2020 | HOOD05162003113055 USA - Effective Date: 11.24.2020 | HOOD05162003144668 To add the Access-i, Dot Add-in: Open the Dot Cockpit Select the Program Editor > Edit subtask card and the program step you want to modify On the right of the screen, select the “Default” Add-in tab, then select “Access-i” and drag the add-in into the program step Dot Cockpit - Program Editor (Program Path = XA30 User Tree/Brain/clinical libraries/general) O x Explorer Program Editor Connection to Fleet database interrupted Browse Import Export Edit Simulate × . Standard Browse Search Default XA30 User Tree >> Brain >> clinical libraries >> general Patient View Addins [12] AAhead_scout 00:19 Default AddIns [39] Access-i Abdomen Bolus Timing t1_se_r_sag 03:25 ABLE t1_se_cor 02:25 Access-i Access-i Advanced Application t2_tse_tra_512 03:37 + Insert Addin Angio Bolus Timing ep2d_diff_4scan_trace_p2 02:49 AutoAlign AutoAlign Scout t2_tse_dark-fluid_tra 04:50 Basic Decision BOLD tof_f13d_tra 05:12 Breast Auto Coverage Scout Breast Basic Repeat this process for each program step that you want to control from the client. You now have the capability to control and edit the program steps from the client. For special interactive sequences, you can adjust the measurement parameters in real time while the measurement is running. To allow the adjustment of these parameters remotely, you need to configure the Access-i Dot add-in that is assigned to the program step: Open the Dot Cockpit and select the Program Editor > Edit subtask card and select the Program Step Right-click and select Edit Properties from the context menu The Step Properties dialog window opens. In the lower-left of the Step Properties dialog window, click AddIn Configuration to open the Dot Add-in configurator dialog window Select the Access-i card and then select Enable Interactivity Click Close to confirm To establish active Access-i Remote Control, these preconditions must be realized: First register a patient Once the patient is registered in the Exam task card, ensure that “No sequences” are “open or running” in the measurement queue In addition, “no pop-up dialog windows can be open” The window “Access-i remote Control Active” will be overlaid on top of the syngo Acquisition workplace informing you that the remote control is now active You are now ready to control the syngo Acquisition Workplace remotely using Access-i. Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 05 MAR 2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 10 SEP 2020 | HOOD05162003113055 USA – Effective Date: 11.24.2020 | HOOD05162003144668 Healthineers Page 3 of 5 Job Aid | Access-i Remote Control Effective Date: 10 SEP 2020 | HOOD05162003113055 USA - Effective Date: 11.24.2020 | HOOD05162003144668 To Cancel the Active Remote Control, select the Stop Scanning button. This cancels both “Active” and “Passive” control. Scanner in Remote Control This device is currently being controlled by Remote Client 1 Therefore no user interaction is possible. Stop Scanning Disconnect Client(s) Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 05 MAR 2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 10 SEP 2020 | HOOD05162003113055 USA – Effective Date: 11.24.2020 | HOOD05162003144668 Healthineers Page 4 of 5 Job Aid | Access-i Remote Control Effective Date: 10 SEP 2020 | HOOD05162003113055 USA - Effective Date: 11.24.2020 | HOOD05162003144668 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively ‘Operator Manual’) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively ‘Functionality’) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 05 MAR 2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 10 SEP 2020 | HOOD05162003113055 USA – Effective Date: 11.24.2020 | HOOD05162003144668 Healthineers Page 5 of 5
- MR
- Access-i
- Remote
- Control