ACUSON S Family Ultrasound System Acquiring an Endovaginal 3D Volume
ACUSON S Family Ultrasound System Acquiring an Endovaginal 3D Volume
ACUSON S Family™ Ultrasound System Acquiring a Endovaginal 3D Volume 1. Select 9EVF4 transducer and appropriate EV preset. ATE LE DEVF Fetal Echo GYN Neo Head OB User Defined FR Control Off Reset Steer Angle Steer 0" L/R Flip Off UID Flip Off Live Dual Off ASC DUAFORT Blopsy Off Features Sie U Virtual Touch 2. Optimize 2D image: Press 4D hard key. 20 Dual DVR 4D M 3. Choose 3D with 3D/4D Soft Key. THI 3D 4D Start 2D FOV 4D ROI Select Position Size SIEMENS www.usa.siemens.com/healthcare 589 Effective Date: 09/06/2016 Page 1 of 3 4. Drop-Down menu should be on Auto-Sweep. 9EVF4 GYN Imag -TEO Set Reset Steer Angle Steer0" Auto Sweep Angle 90" Quality 6 3D ROI On Features Sim El Virtual Touch Protocol 5. Choose widest Angle sweep (90 degree – Quality 6) for the uterus. :XIFEN? 9EVF4 GYN Image 3D TEO Set Reset Steer Angle Steer0 Auto Sweep Angle so- di Quality 5 3D ROI On Features A Sin El Virtual Touch Protocol 6. Make sure 3D ROI is ON and make as large as possible. 7. Make sure the entire fundus of the uterus is in the FOV. SIEMENS www.usa.siemens.com/healthcare Page 2 of 3 8. Press Clip Store to save the volume sweep. Clip Store 9. You can now manipulate the volume. SIEMENS 9EVF GYN Genera 2D -- 100% GE 9.00 MHZ 4dB DR70 SC DTCE Of MapD/ST SIEMENS www.usa.siemens.com/healthcare Page 3 of 3