ACUSON SC2000™ Volume Imaging Ultrasound System – Scanning with eSieScan Protocols
This is a step by step job aid that reviews the process of using an eSieScan protocol while scanning a patient
ACUSON SC2000™ Volume Imaging Ultrasound System – Scanning with eSieScan Protocols Ultrasound: Cardiology 1. Go to the patient demographic page via the new patient key. 2. Click on the drop down menu next to workflow. S/N: 400631 Version: USD_ETCLATEST_SC2000_3.0_INTEG49_RELEASE Date: Monday, June 10, 2013 / 07:27 - PaCers Demographica Last Name First Name Date of Birth Mode Name Age Sex Misle . Request information Refernig Physicon Zit.Don Nane SEMENS HEALTHCARE Accession Number Partanning Physician Request D Operator . Workflow Exam information Trieso Exam Preset Cardiac Hecha Weg's Additional Iortracion BSA - Curtos Information Custom 1 Custom 2 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 SIEMENS Master Template: EFF Date 22 Aug 17 | HOOD/HILS HOOD05162002977513 Healthineers Jobaid: Effective Date: 12/07/2018 | HOOD/HILS Job Aid | ACUSON SC2000™ Ultrasound System - Scanning with eSieScan Protocols Effective Date: 12/07/2018 | HOOD05162002977513 3. Select the eSieScan protocol you would like to use. Report and Worksheet Patient ID Date of Birth MM/dd/yyyy Age Sex Male Female . Other Institution Name Performing Physician Operato Workflow Cardiac 14 BP (s/ mmHg Indication Additional Information 4. If not selected from the demographic page, click on the Workflow menu in the upper left corner during live imaging to select a workflow protocol. Select protocol from the drop down menu A protocol can be selected and started at any point in an exam SIEMENS Cardiac Cardiac Adult Echo Protocol Acquisinon Limited Adult Echo 2D Exercise Stress 2-Stage Transducer Vascular ECHO CF RETRO 2 2D Feraittines Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 2 of 7 Job Aid | ACUSON SC2000™ Ultrasound System - Scanning with eSieScan Protocols Effective Date: 12/07/2018 | HOOD05162002977513 5. Once you select a protocol and begin imaging, the protocol will load on the right side of the screen. SIN: 400631 Version: USO_ETCLATEST_SC2000_10_INTEGIS_RELEASE Date: Friday, June 07, 2013 / 17:17 AS Eche Protocol PROTOCOL 130807044744484 06/07/2013 5:17:16 PM . 098 / ME: 1.23 / TIS: 1.20 New Step / Cardac / 4Vie Atvit Etha Protocol 5 104 / 130 mm NTHI General 20 1572 / 1013 A -- 6. To start the protocol, click on resume protocol. You may pause and resume the protocol as much as you would like during the exam 4Vic 61 fps / 160. ECHO CF' / NTHI General MY ECHO PROTOCOL 14.25 MHz . PLAXOO) -6 dB / DR: 65 dB PLAX COLOR(D.C) LV M-MODEQD.M) , PVPWIRD.C.PW) LV SWEEPQ0) LAS BUBBLE STUDYL. LV VOLUME EF (40) Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 3 of 7 Job Aid | ACUSON SC2000™ Ultrasound System - Scanning with eSieScan Protocols Effective Date: 12/07/2018 | HOOD05162002977513 7. A green triangle will appear next to the step name when you begin. After you complete a step, there will be a check mark next to it and the green arrow will move to the next step. AVIC 1 fps / 160 mm MY ECHO PROTOCOL ECHO CF' / NTHI General 144 25 MHLE - PLAYOOI 11 CD / DR: 65 CD RVD & PLAX 23 / 225 100 13 E PLAx LVI & PLAY LVPW &. PLAX Store Meurge MIL PLAX LVO & PLAY LYPR E. PLAX PLAX COLORDOC) LYDUODEGOM PVPWODC.PM LVENCEPCO MAS BUBBLE STUDY! LV VOLUME EFSACO 2,36 cm 24 cm 4.34 cm 71 bpm 8. To redo a step, press the step name with the onscreen pointer. The picture you originally ac- quired will be saved along with the new one. 4Y1c 61 fps / 160 mm ECHO CF' / NTHI General MY ECHO PROTOCOL -2D 14.25 MHZ AX(20) ·6 dB / DR: 65 dB PLAX COLOROO.C) LVM-MODEQD.M) PVPWOD.C.PW) LV SWEEPQO) IAS BUBBLE STUDYL LV VOLUME EF(AD) 9. To skip a step, click on the next desired step with the pointer. The green arrow will move for- ward and the skipped steps will not have a check mark. Adult Echo Protocol Pause Prefer al "PLAX LVCD) PLAX LVCD) PLAX LVRD.C) PLAX AOVILARD,M) Store Image PLAX MVRD.M) PLAX LVRD.M) PLAX TV(20) PLAX TVQD.C) PBAX MV PEAX PAP PSAX APEX PBAX PAPCD.C) PRAX AVRO) PSAX AV(O.C) PBAX PVCD.C) . PRAX PVRD.C.CW) PSAX TVQ2D.C) PBAX TV(2D.C.CW) MACRO) A4C MVQD.C) MAC MVGO.C.PW) MAC MV(D.C.CW) DTI LAT DTI SEPTAL ASC AoVQDI End Protocol Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 4 of 7 Job Aid | ACUSON SC2000™ Ultrasound System - Scanning with eSieScan Protocols Effective Date: 12/07/2018 | HOOD05162002977513 10. To skip a measurement within a step, press the delete set soft key option. A red X will appear next to the skipped measurement. 19 598 / 160 mm MIT ECHO PROTOCOL ECHO CF"/ Cerenal 4Vlc 19 fps / 160 mm MY ECHO PROTOCOL H4 25 MHZ ECHO CF / General -Color COV/20ME H4.25 M-te . PLAXOO) PROWIEDIS EN 11 GB / DR: 65 CB VI PLAX COLORGO,C) PRI C/d # 29 mmH 5 mes / 1.75 MHr -Color- CDV / 2.0 MHZ VILVI/-MODECOM -0.5 CB PVPW20.C./W) 93 BUBBLE STLEIL PW- 3.5 mm / 1.75 Mitz 14 GB / DR: 70 PRed Vel Time (sec): 10 Sitoes Image 100 LV SHEEPDO) IAS BUBBLE STUDYT LV VOLUME EFIZO) 11. Press the Review Key to review images. To delete an image, press the next key (right click) on image and select delete. Review Cadence Dual Dyn R MultiHz Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 5 of 7 Job Aid | ACUSON SC2000™ Ultrasound System - Scanning with eSieScan Protocols Effective Date: 12/07/2018 | HOOD05162002977513 12. The protocol will automatically pause when the following occurs: You go into review mode You initiate a mode or measurement not in the protocol The heading will read Resume Protocol SN: 400631 Version: USO ETCLATEST 5C2000_10_INTEGER_RELEASE Date: Friday, June 97, 2013 / 17:13 130607044746404 20 13. Once the protocol is finished, you may continue scanning or press end protocol 4VIC fps / 160 min MY ECHO PROTOCOL CHO CF /NTH General 14.25 Mitz . PLAXIOI 11 CD / DR: 65 co 1607 / 1633 PLAX COLOROD.C) LVMJIODEGOM LV SWEEPGO IAS BUBBLE STUDYE LY VOLUME EFID 14. Once you have ended a protocol, you will not be able to restart the protocol but you will be able to continue capturing more images If you want to repeat the same protocol during the exam, you have to enter a different protocol first (see step 4) and then go back and choose the original protocol again. o The origningal protocol will load and all check marks will be removed since the system see it like you are scanning with that protocol for the first time o No images with be lost o All measurements from entire exam will still be averaged Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 6 of 7 Job Aid | ACUSON SC2000™ Volume Imaging Ultrasound System - Creating a New eSieScan Workflow Protocol Effective Date: 08/16/2016 | Quality HOOD05162002837013 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the lates t version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 7 of 7
- ACUSON SC2000™ Volume Imaging Ultrasound System – Scanning with eSieScan Protocols