Replacing the Hemoglobin Lamp Video
This video provides instructions for replacing the Hemoglobin Lamp on the ADVIA® 2120i Hematology System. The hemoglobin lamp must be adjusted after being replaced. Refer to the Adjusting the Hemoglobin Lamp video.
- hemaglobin
- hemoglobing
- Hgb
- turn off instrument
- open doors
- wait 10 minutes
- phillips head screwdriver
- white cap
- outside of UFC
- remove cap
- disconnect cables
- loosen set screw
- turn one quarter to one half turn
- remove
- tighten set screw
- reconnect cables
- replace cap
- close doors
- turn instrument on