PEPconnect

Removing the Reagent Probes Video

This video provides instructions on how to remove the reagent probes on the ADVIA Centaur® XP system. 

Turn mechanics off and remove the reagent probe cover. Lift the reagent probe tubing. And use pliers to loosen the reagent probe sleeve. Turn the sleeve clockwise and pull the probe and sleeve down into the rent station. Remove the probe and sleeve from the system. Place a new reagent probe in the sleeve. And return it to the system. Lift the probe up into the heater coil and turn it counterclockwise. Tighten the sleeve 1/4 turn with pliers and connect the tubing to the top of the probe.

  • troubleshooting
  • common errors
  • centauro
  • centuar
  • centuarxp
  • centaurxp
  • centauroxp
  • centuarxp
  • cenatur
  • cenaturxp
  • cen
  • cenxp
  • centar
  • centarxp
  • centau
  • centauer
  • centauerxp
  • centure
  • centurexp