Cleaning the Ancillary Reagent Probe Video
This video provides instructions on how to properly clean the ancillary reagent probe on the ADVIA Centaur® XPT system.
ancillary reagent probe after the mechanics have been turned off, use the key to unlock the top cover and then lift the cover open. Move the in process Q to the right. Pull the probe tordue by holding the top of the probe assembly. Check the probe for damage and to ensure it's not bent. Use a gentle downward motion to wipe the probe with lint, free tissue or gauze saturated with cleaning solution. Thoroughly rinsed the probe with lint free tissues or gauze saturated with reagent water. Close the instrument cover and turn system mechanics on.
ADVIA Centaur XPT System Ce XPT System Clean Ancillary Reagent Probe
- monthly maintenance