Cleaning the Aspirate Probes Video
This video provides instructions on how to properly clean the aspirate probes on the ADVIA Centaur® XPT system.
aspirate probes after the mechanics have been turned off, use the key to unlock the top cover and then lift the cover open. Move the in process Q to the right. Carefully lift the probe arms up to the highest position. It's not necessary to remove the aspirate probes. Clean the visible portion of each probe using a gentle downward motion with lint free tissues or gauze saturated with cleaning solution. Rinse each probe using a gentle downward motion with lint free tissues or gauze saturated with reagent water. Close the instrument cover and turn system mechanics on.
ADVIA Centaur XPT System XPT System Clean Aspirate Probes
- monthly maintenance