Cleaning the Cuvette Chute Video
This video provides instructions on how to properly clean the cuvette chute on the ADVIA Centaur® XPT system.
cuvet waste shoot, first open the cuvet waste area door and then remove empty sample tip trays and covers from the Tiptree. Been lift the Tiptree been up and then out of the system remove the cuvet waste bin, empty it and set it aside. Remove the cuvet waste chute by releasing it from the magnetic fitting. Slightly rotate the cube at waist, shoot, then pull it out and down. Soak the cuvette waste chewed and cleaning solution for five minutes while the qubit waste chute is soaking clean the cuvet waste reservoir with lint free tissue or gauze saturated with cleaning solution and then rinse the reservoir with lint, free tissue or gauze saturated with reagent water. When the cuvet waste chute has finished soaking, clean the shoot with the large maintenance brush. An cleaning solution. Rinse the shoot with water, dry the inside of the chute with paper towels. Install the cuvette waste chute by placing the locator pins at the bottom of the shoot in the reservoir. Lift the top of the shoot until it attaches to the magnetic fitting. Ensure that no wires or cables are interfering with the correct positioning of the cuvette waste chute. Install the sample Tiptree been. Install the cuvette waste bin and close the door. If you emptied the cuvette waste bin or the Tiptree been respond to the onscreen prompts accordingly.
ADVIA Centaur XPT System Clean Cuvette Chute CIL 000 22 10708697
- as needed maintenance
- centauro
- centuar
- centuarxpt
- centaurxpt
- centauroxpt
- centuarxpt
- cenatur
- cenaturxpt
- cen
- cenxpt
- centar
- centarxpt
- centau
- centauer
- centauerxpt
- centure
- centurexpt