Cleaning the Primary Reagent Probes Video
This video provides instructions on how to properly clean the primary reagent probes on the ADVIA Centaur® XPT system.
clean the primary reagent probes after the mechanics have been turned off, use the key to unlock the top cover and then lift the cover open. Remove the reagent probes cover. Push the reagent probe assemblies. Tored the incubation ring by pushing the Motors. Use a gentle downward motion to wipe each probe with lint, free tissue or gauze saturated with cleaning solution. Thoroughly rinse each probe with lint, free tissue or gauze saturated with reagent water. Install the reagent probes. Cover and close the instrument cover. Finally, turn system mechanics on.
ADVIA Centaur XPT System AD IA XPT System Clean Primary Reagent Probes y Reagent Probes
- monthly maintenance
- centauro
- centuar
- centuarxpt
- centaurxpt
- centauroxpt
- centuarxpt
- cenatur
- cenaturxpt
- cen
- cenxpt
- centar
- centarxpt
- centau
- centauer
- centauerxpt
- centure
- centurexpt