Cleaning the Sample Tip Chute Video
This video provides instructions on how to properly clean the sample tip chute on the ADVIA Centaur® XPT system.
sample tip waste chute after turning the mechanics off, open the sample tip waste area door. Remove the sample tip waste bin, empty it and set it aside. Pull out the drawer, lift the lid to the cleaning solution bottle and ensure that it remains in the up position. Wipe the cleaning solution tubing with fresh lint, free tissues, or gauze to remove excess cleaning solution from the outside and tip of the tubing. Remove the cleaning solution bottle. Remove the sample tip waste chute. Soak the sample tip waste chute in cleaning solution for five minutes while the sample tip waste chute is soaking, wipe the sample tip waste reservoir with lint, free tissue or gauze saturated with cleaning solution, and then wipe the reservoir with lint free tissues or gauze saturated with water. Dry the sample tip waste reservoir with lint free tissues or gauze. When the sample tip waste Chute has finished soaking, use the large maintenance brush to clean the inside of the chute. Rinse the sample tip waste shoot with water thoroughly. Dry the sample tip away, shoot. Install the sample tip waste chute with the bottom of the chute in the reservoir. Ensure the wires do not interfere with the sample tip waste chute. Install the cleaning solution bottle. Install the sample tip wastebin, close the sample tip waist area door. Turn system mechanics on.
Sli ADVIA Centaur XPT System Clean Sample TIP Chute
- as needed maintenance
- centauro
- centuar
- centuarxpt
- centaurxpt
- centauroxpt
- centuarxpt
- cenatur
- cenaturxpt
- cen
- cenxpt
- centar
- centarxpt
- centau
- centauer
- centauerxpt
- centure
- centurexpt