Monthly Maintenance Job Aid
This job aid provides instructions on how to perform monthly maintenance on the ADVIA Centaur® XPT system.
Monthly Maintenance Cleaning the Primary Reagent Probes 1. Turn the system mechanics off, and then use the key to unlock the top cover. Lift open the cover and the reagent probes cover 2. To clean the primary reagent probes, push the reagent probe assemblies toward the incubation ring by pushing the motors 3. Wipe each probe with lint-free tissue with cleaning solution, followed by thoroughly rinsing each probe with lint-free tissue with reagent water 4. Reinstall the reagent probes cover, close the top cover, and turn the system mechanics back on Monthly HILS340 Effective date 09/04/2014 Maintenance Cleaning the Ancillary Probe and Aspirate Probes 1. Turn the system mechanics off, and then use the key to unlock the top cover. Lift open the top cover 2. To clean the ancillary reagent probe, move the inprocess queue to the right and then pull the probe toward you by holding the top of the probe assembly 3. Clean the probe using a gentle downward motion with lint-free tissue with cleaning solution, followed by thoroughly rinsing the probe with lint-free tissue with reagent water 4. Clean the exterior of the aspirate probes with a lint-free tissue dampened with cleaning solution, followed by rinsing the exterior of the probes with DI water and a lint-free tissue 5. Close the top cover and turn the system mechanics back on HILS340 Effective date 09/04/2014 Cleaning the Air Filter 1. Remove the air filter by lifting the cover up and then pulling the cover out of the bracket 2. Remove the old filter and then install the spare, dry air filter 3. Put the cover back onto the system by pushing in and down at the same time 4. Rinse the old air filter with water and then allow it to thoroughly dry HILS340 Effective date 09/04/2014
- reagent probes
- ancillary probes
- aspirate probes
- air filter
- centauro
- centuar
- centuarxpt
- centaurxpt
- centauroxpt
- centuarxpt
- cenatur
- cenaturxpt
- cen
- cenxpt
- centar
- centarxpt
- centau
- centauer
- centauerxpt
- centure
- centurexpt