Replacing Aspirate Probes Video
This video provides instructions on how to properly replace the aspirate probes on the ADVIA Centaur® XPT system.
one of the aspirate probes after the mechanics have been turned off, use the key to unlock the top cover and then lift the cover open. Then move the in process Q to the right. Locate the probe that will be replaced. Lower the adjacent probe to improve access. Disconnect the tubing on the pro being replaced. If the grounded aspirate probe is being replaced, the grounding wire would also have to be carefully removed from the probe, then push the probe pin down and turn counterclockwise to release the spring loaded probes remove and discard the aspirate probe. Install the new probe. Ensure the probe is seated in the probe guide. Push the probe down and turn the pin clockwise to secure the spring loaded probe. Connect the probe tubing. Raise the adjacent pro back up. Then close and secure the top cover and turn the mechanics on.