Replacing Reagent Probes Video
This video provides instructions on how to properly replace the reagent probes on the ADVIA Centaur® XPT system.
the reagent probes after the mechanics have been turned off, use the key to unlock the top cover and then lift the cover open. And then remove the reagent probes cover. Lift off the reagent probe tubing. And use pliers to loosen the reagent probe sleeve. Turn the sleeve clockwise and pull the probe and the sleeve down into the rent station. Remove the probe and sleeve from the system. Place a new reagent probe in the sleeve and return it to the system. Lift the probe up into the heater coil and turn it counterclockwise. Tighten the sleeve 1/4 turn with pliers and connect the tubing to the top of the probe.