ADVIA® Chemistry XPT System Cleaning Cuvette Detergent Bottle Video
This video provides instructions on how to clean the cuvette detergent bottle. This is a required monthly maintenance task on the ADVIA® Chemistry XPT System.
cuvet detergent bottle may be cleaned when it's refilled, but must be cleaned at least once a month to clean an replenished. The cuvet detergent bottle first opened both fluids doors. Pull the fluid Troy out to gain access to the cuvet detergent bottle. Unscrew the level sensor and filter tubing cap and remove the tubing with the filter and the level sensor. Make sure the tubing with the filter and the level sensor is completely removed from the bottle prior to removing the bottle. Set the filter tubing and level sensor on a lint free cloth near the bottle. Make sure to not disconnect the black level sensor connector. Remove the bottle. Empty any remaining contents of the bottle. Rinse the bottle with the ionized water and drain it well. Refill the bottle with Cuvette detergent solution through the refill cap. Return the bottle to the same position on the fluid tray in the cabinet. Make sure to place the bottle in the correct color coded location. Reinstall the cuvet detergent bottle level sensor and filter tubing cap by turning it clockwise. Return the fluid tray to its original location. Close both fluids doors. Prime the fluid lines by selecting system operations. Then prime. Then Prime 2. Enter 10 or more for the number of times in all fields. Select perform. And then OK, if you're cleaning other bottles, you may wait to perform this priming step for all fluid lines.
SIEMENS SIEMENS ADVIA@ Chemistry XP T System SIEMENS Cleaning and Replenishing the Cuvette Detergent Bottle
- chem
- maint