ADVIA® Chemistry XPT System Cleaning Exterior Panels / Sample Tray Area Video
This video provides instructions on how to clean the exterior analyzer panels and the sample tray area. This is a required monthly maintenance task on the ADVIA® Chemistry XPT System.
Cleaning the exterior analyzer panels and sample tray area must be performed at least once a month. Start the procedure by putting the system in standby mode. Close the system covers panels and doors. Dampen a lint free cloth with one of the prepared solutions and wipe. The exterior surface is. Use a solution of either 10% bleach and water or 70% isopropal alcohol. Using D ionized water and a new lint free cloth, wipe all of the exteriors again. Open the top left cover. Dab and lint free cloth with bleach or alcohol and wipe the exterior surface is of the DP and SPP noise reductions covers. Then used ionized water an A new lint free cloth to wipe the DP and SPP noise reductions covers again. Complete the procedure by closing the top left cover.
ADVIA@ Chemistry XP T System SIEMENS Cleaning Exterior Analyzer Panels and Sample Tray Area SIEMENS