ADVIA® Chemistry XPT System Cleaning Saline Bottle Video
This video provides instructions on how to clean the saline bottle. This is a required monthly maintenance task on the ADVIA® Chemistry XPT System.
Cleaning the Saline diluent bottle can be performed at time of bottle refill, but must be performed at least once a month to clean the Saline diluent bottle. Open both fluids doors. Pull the fluid tray out to gain access to the Saline diluent bottle. Unscrew the level sensor and filter tubing cap and remove the tubing with the filter and the level sensor. Make sure the tubing with the filter and the level sensor is completely removed from the bottle prior to removing the bottle. Set the filter tubing and level sensor on a lint free cloth near the bottle. Make sure to not disconnect the black level sensor connector. Remove the bottle. Empty any remaining contents of the bottle. Rinse the bottle with the ionized water and drain it well. Refill the bottle with 0.9% Saline diluent through the refill cap. Return the bottle to the same position on the fluid tray in the cabinet. Make sure to place the bottle in the correct color coded location. Reinstall the Saline diluent bottle level sensor and filter tubing cap by turning it clockwise. Return the fluid tray to its original location. Close both fluids doors. Nextprime the fluid lines by first selecting system operations, then prime. Then Prime 2. Enter 10 or more for the number of times in all fields. Select perform. And then OK, if you're cleaning other bottles, you may wait to perform this priming step for all fluid lines.
SIEMENS ADVIA@ Chemistry XP T System SIEMENS Cleaning and Replenishing the Saline Diluent Bottle coet
- chem
- maint