ADVIA® Chemistry XPT System Cleaning Wash Solution Reagent Containers Video
This video provides instructions on how to replace or clean the wash solution reagent containers in the RTTs. This is a required monthly maintenance task on the ADVIA® Chemistry XPT System.
clean or replace the wash solution, reagent containers, first open the top right cover. Remove the art covers and set them aside. The covers can be placed on the bracket on the side of the instrument. Remove the wash solution reagent containers from RDT one and RT2. Refer to the online operator guide for the positions of awash containers. Replace the containers with new ones or clean the old containers with deionized water. Refill the containers with fresh solutions. Make sure to put the correct solutions in each container. Refer to the online operator guide For more information. Return the art covers to their original locations. Complete the procedure by closing the top right cover.