Clean Turntable Interiors
This video provides instructions on how to clean the turntable interiors on the ADVIA® Chemistry XPT system.
interiors require cleaning as part of monthly maintenance. The add via chemistry XPT system will alert the operator to pending maintenance needs. The maintenance procedure for cleaning the turntable interiors can be accessed by selecting the topic from the maintenance schedule and selecting the procedure details button. Turn off the system. To access the turntable interiors, begin by opening the DP noise reduction cover. Remove the CT and STT evaporation covers and place on a clean surface. Squeeze the prongs on the DP drip tray. Remove and place on a clean surface. Pull up on the two black fasteners to release the skytrail odor. Using the metal handles, lift out the skytrail odor and place on a clean surface. Pull up on the two black fasteners to release the STT tray loader. Lift out the STT tray loader by the two metal handles and place the STT tray loader on a clean surface. Using lint free cloth, wipe the interior of the see TT&STT housing. Line up the black arrows on the see T and STT tray loaders with the guide pins. Ensure the tray loaders are securely in position. Then push the fasteners in place. Reinstall the DP drip by lining up the tab with the guide pin and pressing firmly into place. Replace the CT and STT evaporation covers. Closed the DP noise reduction cover and finally closed the system cover. Restart the system and document on the maintenance schedule.