Monthly Maintenance Job Aid
This job aid provides instructions on how to perform monthly maintenance on the ADVIA® Chemistry XPT system.
Monthly Maintenance Monthly Maintenance SIEMENS AMAB GERMANY XP de CAUTION 00E 1 READY (OK to replace Samples ce Reagents.) 80 1. Select Maintenance > Schedule to locate the CORESSon Dugrantor Setup monthly maintenance tasks and maintenance Schedule Comme Burk ISE Operation procedures Querdue: 0| Fales: 0 2 • Clean exterior analyzer panels and sample firoutecy Completed tray area Corvette Back Measurement . Perfore Proceder - 41/2014 12:00:00 AUN • Clean or replace RTT wash solution reagent Cuvette Debergent Botte - Des containers 6/1/2014 12:00:00 AM 4/112014 12:50:00 AUN • Clean saline bottle Disitioe Toy Cevettes : Orat Every 1 mes 6/1/2014 12:00:00 208 • Clean cuvette detergent bottle Crverso Condi Sator Borse - Com /1/2014 12:20:00 A8 • Clean LWP line filter ISE Decrode Aisendly Deteryest Welt . Perform 4/1/2014 12:00:00 A • Drain Vacuum Tank and Degasser Ancillary Enagent Botte Filters - De 6/1/2014 12:50:00 AN • Inspect, and clean if needed, the dilution the me Time Fraktion and Divison Crvelles . Replace Every 1 months bowl and waste drain nozzle Onboard Pare /Wat ne Boite TISbe - Clean 6/1/2014-12:00:00 AIN 2. Select the Completed details button to log each ABSALONRY completed maintenance task ? Thewidm May 22, 20 14 2 08 37 AM Note: Navigate to the Maintenance > Log tab to review completed maintenance tasks. Comments can be added to activities and the log can be printed or exported. HILS00001312 Effective Date: 12/22/2014 Monthly Maintenance – Inspect, and clean if needed, the dilution bowl and waste drain nozzle Note: Refer to the ADVIA Chemistry XPT Online Help Guide for full instructions. 1. Inspect the dilution bowl and waste drain nozzle for debris. Clean the dilution bowl with distilled water and waste drain nozzle with a cotton stick, if needed. Monthly Maintenance – Clean or Replace RTT Wash Solution Reagent Containers Note: Refer to the ADVIA Chemistry XPT Online Help GLUN Guide for full instructions. loooo 1. Remove wash solution reagent containers from RTT1 and RTT2 2. Replace the containers with new ones or clean the old containers with deionized water 3. Refill the containers with fresh solutions as specified in the table above 4. Reinsert containers and close all covers HILS00001312 Effective Date: 12/22/2014 Monthly Maintenance – Clean and Replace Saline Dilution Bottle and Cuvette Detergent Bottle Note: Refer to the ADVIA Chemistry XPT Online Help Guide for full instructions. 1. Open fluids doors 2. Pull out fluids tray 3. Unscrew filter cap and remove the tube with filter 4. Remove the bottle from the fluid tray 5. Empty the remaining contents - CD --- 081 6. Rinse with deionized water and drain well 7. Refill the bottle and return to the fluid tray - -- Cuvette Detergent Incubation Bath Oil Cuvette Conditioner Isotonic Saline Dilsen 8. Replace the refill cap and filter cap ----- == 9. Return fluids tray to position 10.Close fluids doors. 11.Prime the fluid lines HILS00001312 Effective Date: 12/22/2014 Monthly Maintenance – Drain Vacuum Tank and Degasser Note: Refer to the ADVIA Chemistry XPT Online Help Guide for full instructions. 1. Open the left panel door to access the blue vacuum tank drain tube and degasser tube on the lower shelf 2. Remove the cap from the waste water drain tube and place the discharge tube in a container 3. Open the drain valves to drain 4. Select Maintenance > Schedule > Drain Vacuum Tank and Degasser 5. Select Perform 1 6. Replace the caps and return the tubes to original location 7. Close the door 8. Close the drain valves 9. Select System Operations > Initialize to return the system to READY HILS00001312 Effective Date: 12/22/2014 Monthly Maintenance – Clean the Large Water Pump Line Filter Note: Refer to the ADVIA Chemistry XPT Online Help Guide for full instructions. 1. Remove the Pure Water Bottle to access the Large Water Pump (LWP) line filter 2. Place a cloth and a container under the filter case to catch residual water 3. Unscrew the filter case to the left and remove 4. Clean the inside of both filter case and mesh filter with a brush 5. Rinse each with deionized water and dry with a lint-free cloth 6. Reinstall the mesh filter inside the filter case and reattach the filter case 7. Return the pure water bottle to its original location 8. Connect all tubes 9. Turn on the LWP 10.Prime the fluid lines 11.Set the water pump pressure, if required, to 68kPa HILS00001312 Effective Date: 12/22/2014
- chem
- maint
- vacuum tank
- degasser
- cuvette detergent
- LWP
- saline bottle
- RTT wash solution