PEPconnect

Agile MAX: Start up to Completion of Fluoroscopy Examinations

This step by step will guide the User through using the Agile MAX system to
from starting the system to performing a Fluoroscopy Exam, saving images and
sending to PACS.

Agile Max: Startup to Completion of Fluoroscopy Examinations XPRF 1. Ensure the FLC is powered on at the exposure console. 2. Select “Update RIS Worklist” from the “Patient” subtask card or by pressing F11 on the key- board. RIS 3. Select the desired patient from the “Pre-Registered Patients” tab by double-clicking on their name. 4. Ensure the necessary Organ Programs (OGPs) are displayed in the box in the upper right cor- ner. Emergency, 14 Patient 799 ESOPHAGRAM T ESOPHAGUS Fluoro_Barium 2fps_BW Fluoro_Barium 2fps_BB Fluoro_Pediatric_Barium 2fps_BW Exar Fluoro Rad Next, is the room set up. If the system has an overhead x-ray tube, place it in the “Park” position (typically in the corner of the room). A large “P” will display on the front of the x-ray tube when the position is reached. Tape DEVICE Measure P Setup ? Emergency, 15 115 T ABDOMEN INFANT Parking pacton Aign tube / 1354 FD 165. 11 Manual .# Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 11/23/2020 | HOOD05162003148552 Healthineers Job Aid | Startup to Completion of Fluoro Exam Effective Date: 11/23/2020 | HOOD05162003148552 5. After placing the patient safely on the table, raise the table to at least the working height of 32 inches. Typically there is a slight bump once the that height is reached. The table can be adjusted higher if desired. If the minimum height is not reached, the following system mes- sage is displayed. i No X-ray. Table at min No DIT (FD) movement - Tube touches the floor. Raise table first OK 6. Press and hold one of the “Unpark” buttons until the DIT is completely across the table. The light over the “Park” button is now a solid green light. The word “READY” should appear in green on the bottom of the smaller monitor in the room (the TUI) as well as in the bottom right corner of the monitor in the control room (the FLC). 7. Images can be acquired. The operator can save a Last Image Hold (LIH) or Store Fluoro Loop during the exam. Note: Fluoro Loop is an option. Las LAST IMAGE HOLD/IMAGE CAPTURE S STORE FLUORO LOOP 8. Once the exam is completed, the DIT will need to be put in the “Park” position. There are four steps required for parking. P a. Press and release one of the “Park” buttons. The green light on it will begin to flash. b. Raise the DIT all the way up until the Vertical Lock light comes on. Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 2 of 4 Job Aid | Startup to Completion of Fluoro Exam Effective Date: 11/23/2020 | HOOD05162003148552 c. Push the DIT transversely towards the wall until the DIT stops and the transverse/longitudinal light comes on. -ED-> d. Press and hold the “Park“ button again until the DIT is completely parked. 9. Images can be reviewed and manipulated (if desired) in the “Post Processing“ tab on the monitor in the control room. 10. To send images to PACS, select the “DICOM Send“ button on the “Documentation“ tab in the “Send“ subtab. Verify all images that are going to PACS have a white dot in the upper left corner. Once sent, images will have the letter “A“ on the top left corner. Note: Right mouse clicking on an image will add or remove the white dot determing rather or not the image will go to PACS. Select Film Send Target 1 Target 2 Target 3 Send Exam Protocol as: SC image DICOM Dose SR Override configured settings ? Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 3 of 4 Job Aid | Startup to Completion of Fluoro Exam Effective Date: 11/23/2020 | HOOD05162003148552 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter co llectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect t he latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written auth ority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 4 of 4