Aptio® Automation Bulk Input Module Maintenance Job Aid
This job aid provides instructions on how to perform weekly and monthly maintenance on the Bulk Input Module on Aptio® Automation.
Note: This job aid is designed for systems which include "AP2" or "APT" and the letter "C" in the serial number. If your system serial number includes "APT" but does not include the letter "C", please select the Aptio Automation (Siemens) training area.
Bulk Input Module Maintenance Aptio® Automation Weekly Maintenance Refer to the Aptio Automation Operations Manual when performing the following procedures. Clean the Worktable Supplies needed: lint-free cloth, 0.1 % sodium hypochlorite solution 1. Bring the Bulk Input Module Off-line: a. Select Overview b. Select Bulk Input in Overview screen c. Select Status d. Select Off‐line, select Do Not Flush 2. Clean the worktable, the lid and the lid button with a lint-free cloth moistened with 0.1% sodium hypochlorite solution. Be careful during the cleaning. Excessive force could damage the lid or the lid button WARNING: Esteate labo lo pof crossed -Sangle top is completely cinese SAMPLE TUDES, PLEASE ENSURE BEFORE LOADING THE HOPPER WITH arcade isbei is completely attached as fast as Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 SIEMENS Master Template Effective Date: 3/11/2019 | HOOD05162002998095 Healthineers Job Aid Effective Date: 8/1/2019 | HOOD05162003026271 Job Aid |Bulk Input Module Maintenance Effective Date: 8/1/2019 | HOOD05162003026271 3. Reset the Bulk Input Module robot: a. Select Bulk Input from the Overview screen b. Select the Diagnostics ‐ Robot c. Select Home d. When robot initialization is completed, select On‐line and confirm Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 2 of 10 Job Aid |Bulk Input Module Maintenance Effective Date: 8/1/2019 | HOOD05162003026271 Monthly Maintenance Clean the robot gripper Supplies needed: Lint-free cloth moistened with 0.1 % sodium hypochlorite solution 1. Ensure the Bulk Input Module is Off-line 2. If the Bulk Input Module robot gripper is closed, perform the following procedure: a. Select Bulk Input in the Overview screen b. Select Diagnostics ‐ Robot c. Select Tube Gripper d. Select Open 3. Access the area over the robot: a. Unscrew the knobs b. Open the cover and, using caution, remove it Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 3 of 10 Job Aid |Bulk Input Module Maintenance Effective Date: 8/1/2019 | HOOD05162003026271 4. Remove all sample tubes, if present 5. Clean the gripper inside and outside with a lint-free cloth moistened with 0.1% sodium hypochlorite solution. Do not allow the cleaning solution to touch any wires, sensors, or other mechanisms. Clean only the inside and outside of gripper jaws 6. Replace the cover and secure Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 4 of 10 Job Aid |Bulk Input Module Maintenance Effective Date: 8/1/2019 | HOOD05162003026271 7. Reset the Bulk Input Module robot: a. Select Bulk Input from the Overview screen b. Select the Diagnostics ‐ Robot c. Select Home d. When robot initialization is completed, select On‐line and confirm Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 5 of 10 Job Aid |Bulk Input Module Maintenance Effective Date: 8/1/2019 | HOOD05162003026271 Clean the tray Supplies needed: Lint-free cloth moistened with 0.1 % sodium hypochlorite solution 1. Bring the Bulk Input Module Off-line: a. Select Overview button b. Select Bulk Input button in Overview screen c. Select Status d. Select Off‐line, select Do Not Flush 2. Remove all sample tubes from the Bulk Input Module tray, if present: a. Select Diagnostics b. Select Diagnostics ‐ Lid, Hopper & Separator c. Select Lid d. Select Open e. Manually remove the sample tubes from the tray 3. Remove the front panel by carefully lifting it with two hands Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 6 of 10 Job Aid |Bulk Input Module Maintenance Effective Date: 8/1/2019 | HOOD05162003026271 4. Remove the tray a. Release the upper panel and lift it up b. Unlock the safety mechanism, pulling it out before trying to pull the tray handle Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 7 of 10 Job Aid |Bulk Input Module Maintenance Effective Date: 8/1/2019 | HOOD05162003026271 c. Pull the handle to allow the tray extraction d. Clean the tray with a lint-free cloth moistened with 0.1% sodium hypochlorite solution Co Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 8 of 10 Job Aid |Bulk Input Module Maintenance Effective Date: 8/1/2019 | HOOD05162003026271 e. Replace the tray and push the safety mechanism to lock it f. Replace the front panel and ensure it is securely in place g. Reset Bulk Input Module Robot: a. Select Bulk Input in the Overview screen b. Select Diagnostics ‐ Robot c. Select Home d. Wait until robot initialization is completed, then select On‐line and confirm Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 9 of 10 Job Aid |Bulk Input Module Maintenance Effective Date: 8/1/2019 | HOOD05162003026271 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Aptio is a registered trademark of Siemens Healthcare Diagnostics. Copyright © Siemens Healthcare GmbH 2019 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 10 of 10
- track
- aption
- automaton
- maintain
- clean
- gripper
- robot