Aptio® Automation Sealer Module Maintenance Job Aid
This job aid provides instructions on how to perform monthly maintenance on the Sealer Module on Aptio® Automation.
Note: This job aid is designed for systems which include "AP2" or "APT" and the letter "C" in the serial number. If your system serial number includes "APT" but does not include the letter "C", please select the Aptio Automation (Siemens) training area.
Sealer Module Maintenance Aptio® Automation Monthly Maintenance Refer to the Aptio Automation Operations Manual when performing the following procedures. Clean the gripper pads Supplies needed: lint‐free cloth, 0.1% sodium hypochlorite solution 1. Bring the Sealer Off‐line and turn off the heater: a. Select Sealer from the Overview screen b. Select Off‐line, select Flush Carriers and confirm c. Select Diagnostics ‐ Sealing Head d. Select Heater and then Off e. Select Refresh to display the information related to the Module 2. Check the heater temperature. Wait until the value corresponding to the Sealer – Heater Analog sensor is 55°C/131°F or less before continuing this procedure. It will take approximately 30 minutes for the value to reach 55°C/131°F Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 SIEMENS Master Template Effective Date: 3/11/2019 | HOOD05162002998095 Healthineers Job Aid Effective Date: 5/16/2019 | HOOD05162003006443 Job Aid | Sealer Module Maintenance Effective Date: 5/16/2019 | HOOD05162003006443 3. Open the Sealer Module cover. Use caution when opening the lid to ensure it is fully open 4. Clean the gripper pads with a lint‐free cloth lightly moistened with 0.1% sodium hypochlorite solution 5. Inspect the gripper pads to ensure they are secure and not excessively worn. If the gripper pads are positioned incorrectly or excessively worn, contact Technical Support Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 2 of 4 Job Aid | Sealer Module Maintenance Effective Date: 5/16/2019 | HOOD05162003006443 6. Using a lint‐free cloth lightly moistened with deionized water, clean the area surrounding the carrier sensor 7. Close the Sealer Module cover 8. Bring the Sealer Module On‐line: a. Select Overview b. Select Sealer c. Select Status d. Select On‐line and confirm Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 3 of 4 Job Aid | Sealer Module Maintenance Effective Date: 5/16/2019 | HOOD05162003006443 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Aptio is a registered trademark of Siemens Healthcare Diagnostics. Copyright © Siemens Healthcare GmbH 2019 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84‐0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 4 of 4
- track
- aption
- automaton
- maintain
- clean
- gripper
- robot