Aptio® Automation (Siemens) Daily Maintenance Job Aid
This job aid provides instructions on how to perform daily maintenance on Aptio® Automation.
Note: This training is designed for systems which include "APT" in the serial number but does not include the letter "C". If your system serial number includes "APT" and the letter "C" or “AP2”, please select the Aptio Automation (Inpeco) training area.
Daily Maintenance Empty Waste • Empty the waste on the Decapper Module • Empty the waste on the Desealer Module • Empty the waste on the Refrigerated Storage Module • Empty the waste on the Aliquoter Module 1. Select module 2. Select Going Off-line 3. Replace the biohazard bags on Decapper, Desealer and Refrigerated Storage Modules 4. Clean Aliquoter Module waste with 10% bleach 5. Bring module on-line Refer to the Aptio™ Automation Operator’s Guide, Preventative Maintenance when performing this procedure. Supplies needed: Biohazard bags ! Observe all warnings and cautions stated in the Aptio™ Automation Operator's guide. 10% bleach Daily Maintenance HILS781 Effective Date: 02/03/2014