Aptio® Automation (Siemens) Troubleshooting Maintenance Job Aid
This job aid provides instructions on how to troubleshoot errors including performing a controlled power shutdown on Aptio® Automation.
Note: This training is designed for systems which include "APT" in the serial number but does not include the letter "C". If your system serial number includes "APT" and the letter "C" or “AP2”, please select the Aptio Automation (Inpeco) training area.
Shut Down and Start Up Shut Down Restart 1. Remove all samples from track 1. Turn on the Main Power Switch 2. Select System on the Overview Screen 2. Wait three minutes before pressing the 3. Select Purge from the function menu I/O Module computer power button 4. Press the check mark in the next dialog box 3. Change login from guest to operator level (100) 5. Wait until all samples have been unloaded 4. Start the Aptio software application 6. Remove all racks from I/O Module 7. All modules/analyzers in error status must be recovered or set to off-line 8. Select Shutdown from the function menu 9. Wait until the application is closed 10.Turn off the Main Power Switch HILS781 Effective Date: 02/03/2014 Troubleshooting Error Recovery SIEMENS Aptio Note: Ensure that the Track Module button is Overview Dimension Vista - Status 08:00:00 not red. If Track Module button is red other Status: On-line - Diag Command: SET-MODE 10^ON- Diag Feedback: OK No Error error recovery procedures will not work. Samples ITEM VALUE Perform Track Module error recover first if Tests Backlog Time (minutes) O needed. Node ID 10 Exceptions Room for Empty Carriers 0 1. Select the module/analyzer in Error Status Room for Routine Samples 0 (red button) Configuration Room for STAT Samples O 2. Confirm next dialog by pressing the System checkmark Counters 3. Select Recovery from the function menu Login Print Start Pause Exit GUI 3 Diagnostics Reset On-line Going Off-line Off-line Reboot Backup Recovery Expected: 0 Incomplete: 0 Complete: 0 HILS781 Effective Date: 02/03/2014 Troubleshooting Error Recovery 1. The Error Explanation Box contains the Error Aliquoter Module Code, the Error Message and the Error Error Code Recover Instructions for the Error or Warning SC015 2. Select Recovery from the Error Explanation Error Message Box Aliquoter Tip Waste Full 3. Select the options that are most appropriate. Error Type If the Recovery is successfully executed, the Red error is automatically reset. Step by Step Recovery Note: Many warning status (yellow) are Possible Causes ull condition. system auto recoverable and require no user e procedure. 1. A Sample Tube is located in the Robot Gripper and it should be intervention. moved to its source (information related to source and destination are displayed here below). 2. A Sample Tube is located in the Robot Gripper and it should be moved to its destination (information related to source and destination are displayed here below). 3. There is no Sample Tube located in Robot Gripper. Source: n/a 2 Destination: n/a Show Help Recovery 3 Choose an Option x 1 2 3 HILS781 Effective Date: 02/03/2014
- track
- aption
- automaton