Archiving QC Images to CD/DVD on the MAMMOMAT Inspiration
Archiving QC Images to CD/DVD on the MAMMOMAT Inspiration 1. Insert CD/DVD into appropriate slot on PC. A message will display on bottom of screen. Medium in CD successfully read 2. In Local Database click on the QC series (Images) to Archive. Local Database ARTIFACT, W_Rt QC-raw [2] SINGLE IMAGES, QC-raw com/ 1111 1 Scheduler Artifact, Mo_Mo [1] SINGLE IMAGES, QC-raw 3. Click on Transfer then Archive to DVD in the drop down. Patient Browser Patient Transfer Edit Vier Import Archive to CD 4. You have to label the CD when used the first time. Must have at least 3 characters Enter Label X Device CD Label on medium Viewing tool OK Cancel Help SIEMENS www.usa.siemens.com/healthcare 701 Effective date 9/8/2016 Page 1 of 2 5. A message will be seen at bottom of screen that it is Recording then finished. .................................... Recording started CD or DVD recording successful 5. Click on Transfer and Eject from CD. Patient Browser Patient Transfer Edit Vie Import Archive to CD Archive to ... Send to ... Export to ... Eject from CD Fiest from SIEMENS www.usa.siemens.com/healthcare Page 2 of 2
- MAMMOMAT Inspiration PRIME Edition