ARTIS icono VE20 Roadmap Job Aid
This PDF job aid will demonstrate the workflow for performing Roadmap on your ARTIS icono VE20.
SIEMENS Healthineers ARTIS icono VE20 Roadmap Template Effective Date: 05/27/2019 | HOOD05162003010284 Job Aid Effective Date: 06/17/2020| HOOD05162003101384 Unrestricted © Siemens Healthineers, 2020 Roadmap SIEMENS Healthineers Table of Contents Introduction 3 ke 2D, 3D mp, sine B .. H Roadmap modes 4 Roadmap workflow steps 5 DSA Roadmap workflow steps 7 Subtracted fluoro workflow steps 10 Roadmap functions via touchscreen 11 2 Effective Date: 06.17.2020| HOOD05162003101384 Unrestricted © Siemens Healthcare GmbH, 2020 Roadmap SIEMENS Healthineers Introduction Roadmap is a special operating mode that allows you to view the region of interest with subtracted fluoroscopy for dynamic navigation during procedures such as angioplasty, catheterization, embolization or plaque removal. Roadmap makes catheter and guide wire positioning easier by emphasizing the vessels and devices. During Roadmap, a vessel map is inverted so that the vessels appear white. The vessel map is overlaid on the live subtracted fluoroscopy image. The vessel map may be created as part of the Roadmap procedure (Roadmap Basic mode) or using DSA Roadmap, you can use a previously acquired DSA scene as a vessel map. Introduction 3 Effective Date: 06.17.2020| HOOD05162003101384 Unrestricted © Siemens Healthcare GmbH, 2020 Roadmap SIEMENS Healthineers Roadmap modes Roadmap basic mode: DSA Roadmap: Subtracted fluoro: In Roadmap (Basic mode) all images including the With DSA Roadmap you can use a previously In Subtracted Fluoro, only one subtraction is vessel map are created during the roadmap acquired DSA scene as the vessel map. performed which emphasizes the devices. workflow. A dedicated subtracted image from the DSA scene Use Subtracted Fluoro when no singularization of must be stored as a reference image and displayed the vessels is required during an examination in the configured reference segment. Roadmap 4 modes Effective Date: 06.17.2020| HOOD05162003101384 Unrestricted © Siemens Healthcare GmbH, 2020 Roadmap SIEMENS Healthineers Workflow steps Select the Roadmap icon: • On the touchscreen select the Roadmap icon in the Roadmap and Overlay ref menu. • If Overlay Reference was activated, it becomes deactivated. Start Roadmap • Press on the fluoroscopy pedal on the footswitch. Vessel phase: Sub-phase 1 • The vessel phase of Roadmap creates the vessel map. • In Sub-Phase 1 the mask images are acquired and a subtracted view is achieved. • Keep your foot pressed on the fluoroscopy pedal, Roadmap will automatically switch to Sub-Phase 2 when the mask image is created. Vessel phase: Sub-Phase 2 • Inject contrast media to fill the vessels of interest. • When you are satisfied with the vessel filling, release the fluoroscopy foot pedal. Roadmap 5 Workflow Effective Date: 06.17.2020| HOOD05162003101384 steps Unrestricted © Siemens Healthcare GmbH, 2020 Roadmap SIEMENS Healthineers Workflow steps Device phase • With the next release of fluoroscopy, Roadmap enters the Device phase. • The Live segment displays the vessel map inverted (white) on the grey subtracted background. • Catheters and guidewires can be positioned using the previously acquired Vessel map for guidance. • Fluoroscopy can be interrupted as often as required. The system remains in the Device Phase and displays additional subtracted images each time the fluoroscopy foot switch pedal is pressed. Roadmap RM Neuro ... In the Roadmap menu you can access further functions related to Roadmap (see page 11), for example: Mode Roadmap Subtr. Fluoro DSA ROMP Reset • Reset the Roadmap mask. Reset A+B Reset A Reset B • Adjust the vessel or catheter contrast, or the anatomical background. Display % x2 Contrast A Zoom A • Adjust the image zoom. Tools CLEARmatch Native Vessel • Activate and de-activate Real-time pixel shift. Progress • Activate and deactivate the vessel map. Show Progr. Dyn. Prog. RDMP Set • Show progress during Roadmap. That is, you can start a new device phase but keep the original vessel map. RM Neuro KO Sys. POS. Roadmap Case Flows Exam Favorite End roadmap • When you would like to return to normal fluoroscopy, deselect the Roadmap icon. Roadmap 6 Workflow Effective Date: 06.17.2020| HOOD05162003101384 steps Unrestricted © Siemens Healthcare GmbH, 2020 DSA Roadmap SIEMENS Healthineers Workflow steps Select/Create the DSA reference image: • DSA Roadmap uses a DSA reference image as the vessel map for Roadmap. • You can use a previously acquired DSA scene or you can acquire a new DSA scene. Select one of the following options to start DSA Roadmap: Option one: Option two: Option three: Select the DSA reference image Start Roadmap and then select the Acquire a DSA scene. before starting Roadmap. DSA reference image. Depending on the acquisition settings: The DSA image is displayed in the After you start Roadmap, if no DSA • A reference image may be stored reference segment. image is displayed in the reference automatically segment, you are prompted by the or - - system to select a reference image. • You must manually store a reference image. DSA Roadmap 7 Workflow Effective Date: 06.17.2020| HOOD05162003101384 steps Unrestricted © Siemens Healthcare GmbH, 2020 DSA Roadmap SIEMENS Healthineers Workflow steps Select the DSA Roadmap icon: • On the touchscreen select the DSA Roadmap icon in the Roadmap and Overlay ref menu. • If Overlay Reference was activated, it becomes deactivated. • If the reference segment does not currently display a DSA image, a message appears on the exam room display prompting you to select a DSA image as a reference image. Move the system to the system position of the DSA reference image: • If required, use the automatic drive to move the system to the same position as the DSA reference image. • Press down on the left joystick and deflect the joystick away from you. • Please note: The automatic table movements possible depend on your table type. Please check the guidance messages on the system. Start Roadmap • Press on the fluoroscopy pedal on the footswitch. Vessel phase: • First, the mask images are acquired and a subtracted view is achieved. • After the mask is set, the display automatically switches to subtracted image display (image – mask). DSA Roadmap 8 Workflow Effective Date: 06.17.2020| HOOD05162003101384 steps Unrestricted © Siemens Healthcare GmbH, 2020 DSA Roadmap SIEMENS Healthineers Workflow steps Device phase • With the next release of fluoroscopy, DSA Roadmap enters the Device phase. • The Live segment displays the vessel map inverted (white) on the grey subtracted background. • Catheters and guidewires can be positioned using the previously acquired Vessel map for guidance. • Fluoroscopy can be interrupted as often as required. The system remains in the Device Phase and displays additional subtracted images each time the fluoroscopy foot switch pedal is pressed. Roadmap RM Neuro ... In the Roadmap menu you can access further functions related to Roadmap (see page 11), for example: Mode Roadmap Subtr. Fluoro DSA ROMP Reset • Reset the Roadmap mask. Reset A+B Reset A Reset B • Adjust the vessel or catheter contrast, or the anatomical background. Display 7% X2 Contrast A Zoom A • Adjust the image zoom. Tools CLEARmatch Native Vessel • Activate and de-activate Real-time pixel shift. Progress • Activate and deactivate the vessel map. Show Progr. Dyn. Prog. • Show progress during Roadmap. That is, you can start a new device phase but keep the original vessel map. RDMP Set RM Neuro KO Sys. POS. Roadmap Case Flows Exam Favorite End DSA roadmap • When you would like to return to normal fluoroscopy, deselect the DSA Roadmap icon. DSA Roadmap 9 Workflow Effective Date: 06.17.2020| HOOD05162003101384 steps Unrestricted © Siemens Healthcare GmbH, 2020 Subtracted fluoro SIEMENS Healthineers Workflow steps Select the Subtracted fluoro icono: • On the touchscreen select the subtracted fluoro icon in the Roadmap and Overlay ref menu. • If Overlay Reference was activated, it becomes deactivated. Start Subtracted fluoro: • Press on the fluoroscopy pedal on the footswitch. • After start of fluoroscopy, normal fluoroscopy images are displayed. • After the mask is set, the display automatically switches to subtracted images. • Fluoroscopy can be interrupted as often as required. • The system remains in the device phase (without showing a vessel tree). • Additional subtracted images are displayed each time the fluoroscopy switch is pressed. End subtracted fluoro • When you would like to return to normal fluoroscopy, deselect the Subtracted fluoroscopy icon. Sub fluoro 10 Workflow Effective Date: 06.17.2020| HOOD05162003101384 steps Unrestricted © Siemens Healthcare GmbH, 2020 Roadmap SIEMENS Healthineers Functions via touchscreen Biplane: Roadmap • Reset Roadmap when motion artifacts arise. Reset • Cancel the vessel map: return to Sub-phase 1. Mode Roadmap Subtr. Fluoro DSA RDMP 1 Reset A+B Set a new mask in both planes 1 2 3 2 Reset Reset A Set a new mask in plane A Reset A+B Reset A Reset B 3 Reset B Set a new mask in plane B 4 5 6 7 Display x2 Display Change of display of the Roadmap scene Contrast A Contrast B Zoom A Zoom B • Change of vessel/catheter contrast or anatomical Tools 4 Contrast A background in plane A. CLEARmatch Native Vessel Change of vessel/catheter contrast or anatomical 5 Contrast B Progress background in plane B. Show Progr. Dyn. Prog. 6 Zoom the image by a factor of 2 and panning in Zoom A plane A. RDMP Set 1. roadmap ... Zoom the image by a factor of 2 and panning in 7 Zoom B plane B. Sys. Pos. Case Flows Display Exam Favorites Roadmap 11 functions Effective Date: 06.17.2020| HOOD05162003101384 touchscreen Unrestricted © Siemens Healthcare GmbH, 2020 Roadmap SIEMENS Healthineers Functions via touchscreen Biplane (continued): Roadmap Tools Change of image processing during Roadmap Mode 8 CLEARmatch Automatic pixelshift during Roadmap Roadmap Subtr. Fluoro DSA RDMP 9 Native display Display of native scene Reset Reset A+B Reset A Reset B 10 Vessel display Display of vessel map Progress Display of progress during Roadmap, i.e. enter a new device phase with a Display /% x2 new device mask but keep the vessel map. Contrast A Contrast B Zoom A Zoom B 8 9 10 11 Show Progr. Progress will be performed once Tools CLEARmatch Native Vessel 12 Dyn. Prog. Progress will be performed with every 11 12 footswitch action without time offset. Progress K Show Progr. Dyn. Prog. 13 RDMP Set Change of Roadmap protocol set. 13 RDMP Set 1. roadmap ... Sys. Pos. Case Flows Display Exam Favorites Roadmap 12 functions Effective Date: 06.17.2020| HOOD05162003101384 touchscreen Unrestricted © Siemens Healthcare GmbH, 2020 Roadmap SIEMENS Healthineers Functions via touchscreen Single plane: Roadmap RM Neuro • Reset Roadmap when motion artifacts arise. Reset • Cancel the vessel map: return to Sub-phase 1. Mode 1 Mask Set a new mask Roadmap Subtr. Fluoro DSA RDMP 1 Display Change of display of the Roadmap scene Reset Reset RDMP Change of vessel/catheter contrast or anatomical 2 3 2 Contrast background. Display 1% x2 Contrast Zoom 3 Zoom/Pan Zoom the image by a factor of 2 and panning. 4 5 6 Tools Tools Change of image processing during Roadmap CLEARmatch Native Vessel 4 CLEARmatch Automatic pixel shift during Roadmap Progress Show Progr. Dyn. Prog. 5 Native Display of native scene RDMP Set RM Neuro 6 Vessel Display of vessel map Sys. Pos. Case Flows Display Exam Favorites Roadmap 13 functions Effective Date: 06.17.2020| HOOD05162003101384 touchscreen Unrestricted © Siemens Healthcare GmbH, 2020 Roadmap SIEMENS Healthineers Functions via touchscreen Single plane (continued): Roadmap RM Neuro Progress Display of progress during Roadmap, i.e. enter a new device phase with a new device mask but keep the vessel map. Mode 7 Show Progr. Progress will be performed once. Roadmap Subtr. Fluoro DSA RDMP 8 Dyn. Prog. Progress will be performed with every footswitch Reset action without time offset. Reset RDMP 9 RDMP Set Change of the Roadmap protocol set. Display % x2 Contrast Zoom Tools CLEARmatch Native Vessel 7 8 Progress Show Progr. Dyn. Prog. 9 RDMP Set RM Neuro Sys. Pos. Case Flows Display Exam Favorites Roadmap 14 functions Effective Date: 06.17.2020| HOOD05162003101384 touchscreen Unrestricted © Siemens Healthcare GmbH, 2020 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. The information in this material contains general technical descriptions of specifications and options as well as standard and optional features that do not always have to be present in individual cases. Certain products, product related claims or functionalities described in the material (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH, 2020 15 Effective Date: 06.17.2020| HOOD05162003101384 Unrestricted © Siemens Healthcare GmbH, 2020
- artis
- artsi
- icono
- artisicono
- angio
- angiography
- interventional
- floor
- biplane
- UI
- user-interface
- user
- interface
- software
- iconoSW
- ARTIS
- icono
- syngo
- application
- software
- SW
- VE20
- roadmap
- subtraction
- overlay
- dsa
- dsaroadmap