ARTIS pheno Daily System Checks/Troubleshooting Job Aid
Daily System Checks/Troubleshooting Job Aid ARTIS pheno 1. Turn System On (1) Make sure the on-site main switch is switched on and the emergency SHUTDOWN button is not active. (2)- (E) 2. Unplug the wireless foot pedal 3. Test Fluoro and Acquisition 4. The imaging system keeps you continuously informed about the status of the system, check system messages on the message bar on the Exam Room/Control room monitors If any message has “SC” on the message bar, contact the Siemens Service Center • 1-800-888-7436, ask for Technical Support, provide your Functional Location # 5. Perform safety checks, before each examination, check that all displays and signal lamps are functioning properly Emergency STOP buttons, confirm all emergency stops are not pushed in STOP System movements Collision sensors Radiation indicators Image orientation, image rotation, collimation FD input format/zoom factor 6. "Activate Stream" to display Siemens Live and Reference image on the 3rd Party monitor as applicable. Confirm Block Movement and Block Radiation buttons are not active 7. View and configure the Siemens Large screen display if applicable Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 SIEMENS Healthineers Master Template: EFF Date 22 Aug 17 | HOOD/HILS HOOD05162002647849 Jobaid: Effective Date: 11/08/2018 | HOOD/HILS HOOD05162002967879 Job Aid | ARTIS pheno Effective Date: 11/07/2018| HOOD/HILS HOOD05162002967879 AT THE END OF THE DAY: 1. Shut System down-OPTIONS