Atellica® CH Analyzer with Sample Handler High-Volume Pretreatment Job Aid Version 1.25 and Below

This job aid describes how to change the high-volume pretreatment bottle on the Atellica® CH Analyzer with Sample Handler.

Atellica® CH Analyzer with Sample Handler SIEMENS High-Volume Pretreatment Job Aid Healthineers Replacing the High-Volume Pretreatment Bottle from Module Display For laboratories processing high volumes of Hemoglobin A1c_E tests, a 180-mL bottle of pretreatment is stored onboard. Each 180-mL bottle is reusable. The laboratory stocks a 1-L bottle of pretreatment then rinses, dries, labels, and fills the 180-mL bottles as required. To prepare the 180-mL pretreatment bottle: 180-mL 1. If appropriate, rinse all empty 180-mL pretreatment bottles and caps with special reagent water and let them completely dry before filling 2. To fill empty 180-mL pretreatment bottle: A. Using the 1-L pretreatment bottle, fill a 180-mL bottle to the fill line with 180-mL of pretreatment B. Place the cap on the prefilled 180-mL pretreatment bottle C. Record the Lot and Expiration on a label and affix it to the bottle NOTE: When filling the 180-mL bottle, be careful to avoid introducing bubbles into the pretreatment. D. Place the cap on the 1-L pretreatment bottle 3. Store the capped opened 1-L pretreatment bottle at 2–8°C until the NOTE: If a bottle has an existing label, ensure the Lot and Expiration are the same as on the 1-L pretreatment bottle. If not, remove the expiration date existing label and replace with a new label that contains the appropriate Lot and Expiration. *This procedure is continued on the next slide. 1 Template Effective Date: 05/27/2019 | HOOD05162003010284 Effective Date: 01-APR-2021 | HOOD05162003178151 Unrestricted © Siemens Healthcare GmbH, 2021 Atellica® CH Analyzer with Sample Handler SIEMENS High-Volume Pretreatment Job Aid Healthineers Replacing the High-Volume Pretreatment Bottle from Module Display SIEMENS CI-CH (continued) ¿ LabManager Standby To replace the 180-mL pretreatment bottle: 3 0 1. Stop the CH analyzer and unlock the front cover: A. At the module display, select Module State Module State System Fluids Supplies IMT Supplies B. Select the Pause icon C. Select the icon to unlock the CH front cover 2. Lift open the CH front cover Reagent Loader 3. At the module display, select Supplies SIEMENS Supplies n 4. Select Manual Entry ¿ LabManager Standby 5. Locate the prefilled 180-mL pretreatment replacement bottle, enter the Pretreatment Lot and Expiration, then select OK Lot --- Lot Expiration --- • If the Lot is already entered into the analyzer, the Expiration OBS Instructions for automatically populates Status --- scanning 6. Set this bottle on a flat surface, such as behind the module display 4 Manual Entry ? *This procedure is continued on the next slide. 2 Effective Date: 01-APR-2021 | HOOD05162003178151 Unrestricted © Siemens Healthcare GmbH, 2021 Atellica® CH Analyzer with Sample Handler SIEMENS High-Volume Pretreatment Job Aid Healthineers Replacing the High-Volume Pretreatment Bottle from Module Display (continued) 7. Locate the onboard pretreatment bottle notched lid 7 8. To remove the empty 180-mL pretreatment bottle: A. Grasp the notched lid then pull up and out of the analyzer B. To remove this notched lid, hold the bottle with 1 hand then grasp the lid with the opposite hand and turn it counterclockwise ET C. Inspect this lid for debris then, if appropriate, rinse with special reagent water and dry with lint-free tissues NOTE: Do not tilt the opened replacement bottle or the empty notch lid bottle of pretreatment. This can cause spills. If spills occur inside the analyzer, B clean it with reagent water and dry all surfaces with a lint-free tissue. NOTE: Do not handle pretreatment solution without wearing protective equipment, including gloves, laboratory coat, and safety glasses or a To protective face shield. Pretreatment solution can cause skin irritation. A *This procedure is continued on the next slide. 3 Effective Date: 01-APR-2021 | HOOD05162003178151 Unrestricted © Siemens Healthcare GmbH, 2021 Atellica® CH Analyzer with Sample Handler SIEMENS High-Volume Pretreatment Job Aid Healthineers Replacing the High-Volume Pretreatment Bottle from Module Display A B (continued) 9. To install the prepared 180-mL pretreatment replacement bottle: A. Place the notched lid removed in step 7 on the prefilled bottle and turn clockwise until tight B. Grasp the notched lid of the bottle then insert the full pretreatment bottle into the keyed slot 10. Close the CH front cover 11. At the module display, select OK 12. Resume the CH analyzer by selecting the Resume icon Pretreatment Scanned X Pretreatment × 13. Rinse the empty pretreatment bottle removed from the analyzer and the Lot 1222222222 Expiration 2/25/2017 12:00:00 AM cap removed from the stored bottle with special reagent water and let them completely dry Place Pretreatment in the holder, close the front cover, and start the analyzer. ? 11 OK Cancel 4 Effective Date: 01-APR-2021 | HOOD05162003178151 Unrestricted © Siemens Healthcare GmbH, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability may vary from country to country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2021 Global Siemens Healthineers Headquarters Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84 0 siemens.com/healthineers 5 Effective Date: 01-APR-2021 | HOOD05162003178151 Unrestricted © Siemens Healthcare GmbH, 2021