Atellica® Data Manager Configure QC Alerts Job Aid
This job aid provides instructions on how to configure QC alerts and customize severity colors within the Atellica® Data Manager software.
Configure QC Alerts Atellica® Data Manager Configure QC Severity Limits for Alerts 1. From the Command bar, select the Menu icon 2. Navigate to Configuration > Instruments > Instruments The Instruments tab opens under the Utilities tab 3. Right click on the appropriate instrument 4. Select Related > Tests Heolthinkers Atellica ® Data Manager VM-TST-ADM- 9X n Home Sample Query Browser Review and Edit Oc Utilities ... Instruments > i Name Parent instrument New Advia120_1 Edit Advia1650_1 View 3 Advia2120_1 Flags Deep copy Delete Instrument groups + Advia70_1 QC close out Instrument racks Select First to Current AdviaCentaurC. Query LAS test availability LAS channel ... Setup work order download Select Current to Last LIS channel QC instruments AdviaCentaur_1 Setup Westgard On end result message Duplicate AdvialMS_1 Setup targeting Parent instrument Setup LAS connection availability Shielded users Advanced AllQueryChild AUT_AIIG Set group Tests 4 Custom Actions Set routing cluster Atlas_1 Work lists Retarget Counted test AUT_AdviaCent Related Coding system Setup QC severity reset check AUT_AllQuery1 Codes Setup QC severity reset time limit Child instruments AUT_AllQuery2 Barcode translation AUT_AllQuery3 Aspects Default container type AUT_AllQuery4 Default sample type AUT_AllQuery5 Default LIS channel On all query SID translation AUT_AllQuery6 Instrument logs AUT_BloodCult. Sample logs On all query sample select Instrument translators Routing cluster QC Populations ysman, Systemmanager Test availability Tuesday, January 1, 2019 - 22:01 * Procedure is continued on the next page SIEMENS Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Master Template: EFF Date 22 Aug 17 | HOOD05162002647849 Healthineers Jobaid: Effective Date: 6/25/2019 | HOOD05162003011651 Job Aid | Configure QC Alerts Effective Date: 6/25/2019 | HOOD05162003011651 5. Double click on the appropriate test 6. Select the Autovalidation tab 7. Enter a desired value in the QC severity limit for alerts field 8. Select a QC alert type from the dropdown list 9. Select OK Method Test: RBC 6 Instrument: Advia2120_1 Duplicate Basic Advanced Autovalidation Westgard LAS Diff Pad Cut QC severity: 0 Copy QC severity on type error: Paste QC severity on invalid reference: 5 Allow on type error Clear Allow on invalid range Advanced Norm severity limit:? Custom Actions Delta norm severity limit: ? QC severity limit: ? Instrument severity limit: ? 7 QC severity limit for alerts: ? QC alert type: None 8 9 OK Cancel Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 2 of 6 Job Aid | Configure QC Alerts Effective Date: 6/25/2019 | HOOD05162003011651 Configure QC Alerts for Multiple Tests 1. From the Command bar, select the Menu icon 2. Navigate to Configuration > Instruments > Instruments The Instruments tab opens under the Utilities tab 3. Right click on the appropriate instrument 4. Select Related > Tests colthinkers Atellica ® Data Manager VM -TST -ADM 9 X Home Sample Query Browser Review and Edit Qc Utilities Instruments x Name Parent instrument New Advia120_1 ? Edit Advia1650_1 2 View Advia2120_1 Flags Delete Deep copy Instrument groups W Advia70_1 ? QC close out Instrument racks Select First to Current AdviaCentaurC ... ? Query LAS test availability LAS channel Setup work order download Select Current to Last LIS channel QC instruments AdviaCentaur_1 ? Setup Westgard On end result message Duplicate Setup targeting AdvialMS_1 ? Parent instrument Setup LAS connection availability Shielded users Advanced AllQueryChild AUT_AllG Set group Tests 4 Custom Actions Set routing cluster Atlas_1 ? Work lists Retarget Counted test AUT_AdviaCent ... ? Related Coding system Setup QC severity reset check AUT_AllQuery1 Codes 12 Setup QC severity reset time limit Child instruments AUT_AlQuery2 Barcode translation AUT_AllQuery3 Aspects Default container type AUT_AllQuery4 Default sample type AUT_AllQuery5 Default LIS channel ? On all query SID translation AUT_AllQuery6 Instrument logs Sample logs AUT_BloodCult ... ? On all query sample select Instrument translators Routing cluster ? QC Populations sysman, Systemmanager Test availability Tuesday, January 1, 2019 - 22:01 5. Using the checkboxes, select the desired Test(s) 6. Right click and select Set QC severity limit for QC alerts Tests OOK Test Patient res .. Control res. QC severity Targeting Auto t ... Orders |Check moving average New PLT O 7 Edit PLTCLM 0 View RBC 0 Delete RBC Select First to Current 5 RBC Triggering QC result Setup work order download 0 Select Current to Last RDV Setup targeting 0 Setup instrument targeting for rerun Duplicate RDV Setup autovalidation 0 Advanced RDV Reset reagent information 0 Related Custom Actions RET Setup moving average 0 RET Setup request group code Accept results for validated requests 6 Reset QC Severity RTC Set QC severity limit for QC alerts JSTHE Set QC alert type Add PMA Exclusion stoL Setup QC severity reset check 0 Setup QC severity reset time limit sf#MON ? sf#NEU ? o staPMN O ? 4 selected Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 3 of 6 Job Aid | Configure QC Alerts Effective Date: 6/25/2019 | HOOD05162003011651 7. Enter a desired value in the QC severity limit for alerts field 8. Select OK Set QC severity limit for QC alerts QC severity limit for alerts: ? 7 8 OK Cancel 9. Right click and select Set QC alert type Tests Test Patient res. Control res QC severity Targeting Auto t ... Orders Check moving average New PLT ? Edit PLTCLM View RBC 12 Delete RBCE ? Select First to Current Triggering QC result RBO Setup work order download O Select Current to Last RDV Setup targeting 0 ? Setup instrument targeting for rerun Duplicate RDV Setup autovalidation Advanced o RDV Reset reagent information ? Related Custom Actions RET Setup moving average 0 ? RET Setup request group code Accept results for validated requests ? Reset QC Severity O RTC Set QC severity limit for QC alerts ? STHE Set QC alert type ? LO Add PMA Exclusion STAL Setup QC severity reset check ? Sf#M Setup QC severity reset time limit 0 ? ##MON ? ? sf#NEU ? ? Sf#PM ? ? ? O 4 selected 10. Select a desired option from the QC alert type dropdown 11. Select OK Set QC alert type QC alert type: None 10 11 OK Cancel Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 4 of 6 Job Aid | Configure QC Alerts Effective Date: 6/25/2019 | HOOD05162003011651 Customize Severity Colors 1. From the Command bar, select the Menu icon 2. Navigate to System Management > Customize > Severity colors configuration 3. Select a Background color from the dropdown menu for the appropriate severity level 4. Select a Font color from the dropdown list for the appropriate severity level 5. Enter Severities Levels for the appropriate severities 6. Select Save Severity Color Configuration Severity Level Low Severity Level Medium Severity Level High Display Color Display Color Display Color Background Color 255 , 255 , 0 Background Color 255, 153, 0 Background Color 255 , 0 , 0 Font Color Dark Font Color Dark Font Color Light Severity Levels Severity Levels Severity Levels Norms Norms Norms 3 N Delta Norms Delta Norms Delta Norms N Qc Qc 21 QC 3 Instrument Instrument 21 Instrument Method Method 2 Method 3 Save Close Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 5 of 6 Job Aid | Configure QC Alerts Effective Date: 6/25/2019 | HOOD05162003011651 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (herein- after collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2019 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 6 of 6
- attelica
- attellica
- atelica
- adm
- atellica dm