Atellica® Data Manager Patient Moving Average Exclusions Job Aid
This job aid provides instructions on how to exclude results from the patient moving average.
Patient Moving Average Exclusions Atellica® Data Manager Exclude Results from the Patient Moving Average per Test 1. From the Command bar, select the Menu icon 2. Navigate to Configuration > Tests > By name The Tests window opens under the Utilities tab 3. Select the desired test(s) to set exclusions for 4. Right click and select Add PMA exclusion Note: The system allows adding only one exclusion type at a time. Healthineers Atellica ® Data Manager VM-TST-ADM - 2 X Home Sample Query Browser Review and Edit Utilities 8| ... Tests Name Ty Description Sample type Seq no New w28 S w28 Serum 10430 Edit w3 S w3 Serum 10230 View w35 S w35 Serum 10440 Delete w4 S w4 Serum 10240 Select First to Current W5 S w5 Serum 10250 Select Current to Last w6 S w6 Serum 10260 Duplicate w7 S w7 Serum 10270 Advanced w8 S w8 Serum 10280 Custom Actions w9 S w9 Serum 10290 3 WBC S White Blond Cell Count Whole blood Setup Westgard 10 + WBCB S Basophil Set group Whole blood 175 WBCP S Peroxid Related * Whole blood Add to LAS test group 170 wrCRP S Wide Ra Set strict targeting m 1572 Add PMA exclusion 4 wx1 S WX1 11390 wx2 S WX2 Serum 11400 wx3 S WX3 Serum 11410 selected sysman, Systemmanager Tuesday, January 1, 2019 - 21:10 * Procedure is continued on the next page Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 SIEMENS Master Template: EFF Date 22 Aug 17 | HOOD05162002647849 Healthineers Jobaid: Effective Date: 6/13/2019 | HOOD05162003011678 Job Aid | Patient Moving Average Exclusions Effective Date: 6/13/2019 | HOOD05162003011678 5. To set a physician based exclusion: a. From the Type dropdown list, select Physician b. Double click in the Physician field and select a Physician ID 6. To set a patient location based exclusion: a. From the Type dropdown list, select Patient Location b. Double click in the Patient location field and select a patient location 7. To set a high or low value based exclusion: a. From the Type dropdown list, select Limit b. Enter a low limit value, if appropriate Enter a high limit value, if appropriate Add PMA exclusion Type: Patient Location Patient location: Physician: ? Low limit: ? High limit: ? Description: OK Cancel 8. Enter a description, if appropriate 9. Select OK Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 2 of 4 Job Aid | Patient Moving Average Exclusions Effective Date: 6/13/2019 | HOOD05162003011678 View All Patient Moving Average Exclusions 1. From the Command bar, select the Menu icon 2. Navigate to Routine > Quality Control > Patient moving average exclusions 3. Enter Test, Instrument and/or Instrument Group values, if appropriate 4. Select OK SIEMENS Atellica ® Data Manager VM-TST-ADM6X olthineers Configuration Tools Routine Order entry Review and Edit ac Utilities ... System management Review and Edit Pa Service Tools Result entry About Atellica Data Manager Patients Exit Physicians Quality control Advia 120 QC Configuration import il Type Test Samples Qc Results Low limit New Reagents Qc Close Out an High limit Qc lots Edit Qc populations Patient moving average exclusions 2 View Delete Patient Moving Average Exclusion query Select First to Current Test: Select Current to Last Instrument: Duplicate Instrument Group: W Advanced 4 OK Cancel Custom Actions sysman, Systemmanager Tuesday, January 1, 2019 - 21:20 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 3 of 4 Job Aid | Patient Moving Average Exclusions Effective Date: 6/13/2019 | HOOD05162003011678 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (herein- after collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2019 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 4 of 4
- attelica
- attellica
- atelica
- adm
- atellica dm