Atellica® Data Manager Reset QC Severity Job Aid
This job aid provides instructions on how to reset QC severity when the assay QC severity of a given QC result exceeds the assay QC severity limit.
Reset QC Severity Atellica® Data Manager Reset QC Severity After the assay QC severity of a given QC result exceeds the assay QC severity limit, the Test, Instrument, and String value columns of all subsequent QC results remain red until the QC problem is addressed and operators perform the Reset Severity function. 1. From the Command bar, select QC 2. Select the Results tab Note: Additionally, the QC Results tab can be accessed from Review and Edit by right clicking on a result and selecting View QC. 3. Select the QC result associated with the assay that needs resetting Operators can reset the severity of multiple assays by selecting multiple QC results 4. Select Reset Severity Healthingers Atellica ® Data Manager VM-TST-ADM- 6 X ÇA 2 Home Sample Query Browser Review and Edit ac Utilities Results Alerts Graph QC Population Status Filter QC Results (Test: PHTN, From: 01/01/2017 00:00:00, To: 02/01/2019 00:00:00) Details Test strument Lot/Species Assessed String value Unit MS WS IS CS C F -3 -2 -1 T 1 2 3 3 Comments PHTN AdviaCentaur_1 40881 14/11/2018 09:42:27 5.35 ug/mL PHTN AdviaCentaur_1 14/11/2018 09:42:27 1.06 Instrument Comments 40881 ug/ml PHTN AdviaCentaur_1 40881 14/11/2018 09:42:27 1.48 va/ml Parse Comments PHTN AdviaCentaur_1 40881 14/11/2018 09:42:27 1.14 ug/ml Failed Westgard Comments PHTN AdviaCentaur_1 40881 14/11/2018 09:42:27 1.09 ug/ml Westgard Severity: 1 Across lot 1-3s, 1-2s PHTN AdviaCentaur_1 40881 27/07/2017 11:45:55 1.22 ug/ml In Int 1 2e 1 0e PHTN AdviaCentaur_1 40881 26/07/2017 09:44:16 1.0 ug/ml User Comments PHTN AdviaCentaur_1 40881 25/07/2017 11:47:06 1.13 ug/ml Flags PHTN AdviaCentaur_1 40881 24/07/2017 07:37:11 1.08 ug / ml PHTN AdviaCentaur_1 40881 23/07/2017 09:56:00 1.04 ug/ml PHTN AdviaCentaur_1 40881 22/07/2017 12:26:35 1.29 ug/ml User Actions N Mean CV% Target value 1.18 Show Graph Show Result Details Current 461 1.24 0.56 Target deviation 0.12 Cumulative 0 Data range 248 1.28 1.65 51.08 4 Omit Add Comment Reset Severity ? sysman, Systemmanager Tuesday, January 1, 2019 - 15:28 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 SIEMENS Master Template: EFF Date 22 Aug 17 | HOOD05162002647849 Healthineers Jobaid: Effective Date: 7/1/2019 | HOOD05162003011687 Job Aid | Reset QC Severity Effective Date: 7/1/2019 | HOOD05162003011687 5. Select the Enable Test After Reset checkbox, if appropriate 6. Enter a comment Note: If the current QC severity is not equal to or higher than the QC severity limit, the system will not store the comment in the database. 7. Select OK Reset QC Severity 5 Enable Test After Reset The following tests will have QC Severity reset and comment appended: Instrument Test Disabled by Reagent Change Advia1650_1 ALB Related Blocked Methods: Instrument Test Reset Related Methods Comment 6 7 OK Cancel 8. Select F5 on your keyboard to refresh the window to display the reset severity change, if appropriate Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 2 of 3 Job Aid | Reset QC Severity Effective Date: 7/1/2019 | HOOD05162003011687 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (herein- after collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2019 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 3 of 3
- attelica
- attellica
- atelica
- adm
- atellica dm