Atellica® IM Analyzer Daily Maintenance PEP to Go Version 1.25 and Below

PEP to Go is an education resource material designed to support competency mastery on-the-job.

SIEMENS Healthineers Atellica® IM Analyzer Daily Maintenance PEP to Go Effective Date: 20-DEC-2020 | HOOD05162003152571 Unrestricted © Siemens Healthineers, 2020 Atellica® IM Analyzer SIEMENS Daily Maintenance Healthineers Daily Maintenance ➢ Perform Automated IM Daily Maintenance ➢ Perform IM Autocheck 2 Effective Date: 20-DEC-2020 | HOOD05162003152571 Unrestricted © Siemens Healthineers, 2020 Atellica® IM Analyzer SIEMENS Daily Maintenance Healthineers Performing Automated IM Daily Maintenance SIEMENS . Healthineers Atellica Solution Version. 1.14,0.2983003 P-P1-03 3 IM Daily Maintenance contains automated maintenance 4 System Samples Inventory Calibration Qc Patient Orders Worklist Maintenance Setup activities. The system performs a series of cleaning and Schedule Report Log IMT Data Restore Software Updates Copy Configuration rinsing operations to decontaminate the analyzer at Filter Module: P1-060 Frequency: All Time: Next 24 hours Performed By: All Module P1-060 Module Activity Scheduled Sun-10 Mon-11 Tue-12 Wed-13 Thu-14 Fri-15 Sat-16 6 Perform .. configured times or on demand. The operator can Frequency All P1-060 P1-06 Cleaning the IM Fan Filter . 6/1/2018 7:00 -- Mark as Completed configure the time this is performed. Time Next 24 hours 1-060 -P1-06 Cleaning the Exterior of the Aspirate Probes 6/1/2018 7:00 -- Not Perform. Performed By All P1-060 I-P1-06 Cleaning the Exterior of the Reagent Probes :6/7/2018 1:00 To perform the Daily Maintenance activity manually: Search Text Enter the activity name P1-060 Procedure Details 4-06 Inspecting and Emptying the IM Water Trap : 6/7/2018 1:00 -- Scheduled by Operator 5 Edit/View .. 1. Ensure the analyzer is in ready or standby state 66 Inspecting and Emptying the IM Dryer :6/11/2018 7:00 Scheduled Automatically P1-060 1-P1-06 IM Daily Maintenance :!!:: 6/13/2018 2:00 Delete Scheduled by Service V V V P1-060 2. Ensure cuvettes and IM Cleaner are loaded on the I-P1-06 IM Autocheck .:: 6/13/2018 3:30 V V Failed P1-060 system A Overdue -P1-06 IM Weekly Maintenance Confirmation 1-060 Inprocess 1-P1-06 Replacing the Aspirate Probe Completed P1-060 3. On the Command bar, select Maintenance -P1-06 Replacing the Aspirate Probe 200 X Not Performed >1-060 P1-06 Replacing the Aspirate Probe With Comments 4. Select the Schedule tab P1-060 I.P1-06 View Achaties Replacing the Aspirate Probe Replacing the Reagent Probe ? 5. Select IM Daily Maintenance The maintenance activity IM Daily Maintenance requires 20 minutes. Perform this activity now? 6. Select Perform ? PEP. LabManager 7. Select Yes 8. Wait for maintenance to complete 7 Yes No Note: IM Daily Maintenance is not required when the system performs IM Weekly Cleaning. Note: If the cleaning procedure stops before it completes, the system logs an error and does not allow 3 processing samples until it completes the cleaning procedure or completes a rinse procedure. Effective Date: 20-DEC-2020 | HOOD05162003152571 Unrestricted © Siemens Healthineers, 2020 Atellica® IM Analyzer SIEMENS Daily Maintenance Healthineers Performing Automated IM Autocheck Activity SIEMENS .: Healthineers Atellica Solution Version 1.14.0.2983003 P-P1-03 K 2 IM Autocheck performs a self diagnostics maintenance System Samples Inventory Calibration ac Patient Orders Worklist Maintenance Setup 3 Schedule Report Log IMT Data Restore Software Updates Copy Configuration when the operator selects it. The IM Analyzer pauses Filter Module: P1-060 Frequency: All Time: Next 24 hours Performed By: All sample processing until the autocheck completes in Module P1-060 5 Module Activity Scheduled Sun-10 Mon-11 Tue-12 Wed-13 Thu-14 Fri-15 Sat-16 Perform .. approximately 10 minutes. The system automatically Frequency All P1-060 I-P1-06 Cleaning the IM Fan Filter . 6/1/2018 7:00 Mark as Completed Time Next 24 hours P1-060 performs the autocheck daily at a specified time or it can -P1-06 Cleaning the Exterior of the Aspirate Probes 6/1/2018 7:00 -- Not Perform .. Performed By All P1-060 I-P1-06 Cleaning the Exterior of the Reagent Probes -- be manually performed on demand. The system alerts the : 6/7/2018 1:00 Search Text Enter the activity name. P1-060 Procedure Details -P1-06 Inspecting and Emptying the IM Water Trap 6/7/2018 1:00 -- operator 15 minutes before the scheduled activity begins. Scheduled by Operator P1-060 Edit/View .. 1-P 1-06 OCS Inspecting and Emptying the IM Dryer : 6/11/2018 7:00 500 Scheduled Automatically 4 P1-060 1-P1-06 IM Daily Maintenance To perform the IM Autocheck activity manually: of Scheduled by Service :::: 6/13/2018 2:00 Delete V V V P1-060 V I-P1-06 IM Autocheck .::: 6/13/2018 3:30 V V Failed V P1-060 1. Ensure the analyzer is in ready or standby state A Overdue 1-P1-06 IM Weekly Maintenance 6/14/2018 5:00 P1-060 Inprocess -P1-06 Replacing the Aspirate Probe 1 015 8/12/2018 8:07 -- 2. On the Command bar, select Maintenance Completed P1-060 -P1-06 Replacing the Aspirate Probe 2 015 8/12/2018 8:07 X Not Performed P1-060 -P1-06 Replacing the Aspirate Probe 3 015 8/12/2018 8:07 -- 3. Select the Schedule tab With Comments P1-060 I-P1-06 View Achatles Replacing the Aspirate Probe 4 015 8/12/2018 8:07 >1-060 I.P1-06 OC Replacing the Reagent Probe 1 015 6/12/2020 8:07 4. Select IM Autocheck Add Activity. Set Warning. 5. Select Perform ? PEP LabManager Wednesday, June 13, 2018 8:07 AM 6. Wait for autocheck to complete 4 Effective Date: 20-DEC-2020 | HOOD05162003152571 Unrestricted © Siemens Healthineers, 2020 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2020 Global Siemens Healthineers Headquarters Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84 0 siemens.com/healthineers 5 Effective Date: 20-DEC-2020 | HOOD05162003152571 Unrestricted © Siemens Healthineers, 2020