Atellica® Solution Restoring Data Job Aid Version 1.25 and Below
This job aid provides instructions on how to restore data on the Atellica® Solution.
Atellica® Solution SIEMENS Restoring Data Job Aid Healthineers The analyzer automatically performs database backups by SIEMENS Healthineers Atellica IM Solution Version 1.13.0.2884004 P-P1-03 default. DOL V 1 System Samples Inventory Calibration Qc Patient Orders Worklist Maintenance Setup 2 Schedule Report Log IMT Data Restore Software Updates This process can be completed while the system processes Copy Configuration • Filter Y Module: All Frequency: All Time: Next 24 hours Performed By: All samples 3 Module All Module Activity Scheduled Fri-09 Sat-10 Sun-11 Mon-12 Tue-13 Wed-14 Thu-16 5 Perform ... Frequency All P1-060 The default settings for automatic maintenance can be I-P1-0 Cleaning the Exterior of the Reagent Probes :3/8/2018 1:00 Mark as Completed • Time Next 24 hours P1-060 -P1-06 IM Weekly Maintenance : 3/8/2018 5:00 -- edited in the Maintenance Schedule screen Not Perform ... Performed By All C-P1-06 C-P1-06 CH Weekly Maintenance : 3/11/2018 4:00 -- V Search Text Enter the activity name. C-P1-06 Procedure Details The operator can also perform database backup manually if C-P1-06 Checking the Lamp Coolant :3/11/2018 6:00 • 4 Edit/View .. Scheduled by Operator Database Backup :!!:: 3/14/2018 0:30 desired Scheduled Automatically P_P1-03 Delete P-P1-03 SH Autocheck of Scheduled by Service :: 3/12/2018 1:00 A VMM VMM Atellica Magline Transport Autocheck .:: 3/12/2018 1:10 V Failed A Overdue IM Daily Maintenance 3/12/2018 2:00 V V Inprocess C-P.1-06 To perform database backup manually: C-P1-06 IMT Daily Cleaning :!::: 3/12/2018 2:00 v V V Completed P1-060 -P1-06 IM Autocheck :: 3/12/2018 3:30 V V V 1. On the Command bar, select Maintenance × Not Performed C-P1-06 C-P1-06 CH Daily Maintenance With Comments 1 3/12/2018 4:00 V v V V C-P1-06 C-P1-06 3/12/2018 5:00 2. Select Schedule Vww Activities Inspecting the Washer Probes V P1-060 I-P1-06 8 Inspecting and Emptying the IM Dryer 3/12/2018 7:00 A 3. In Module, select All from the drop-down menu Add Activity ... Set Warning ? PEP LabManager Monday, March 12, 2018 11:32 AM 4. Select the Database Backup activity 5. Select Perform... > Yes 6. When Database Backup displays, confirm database backup location or select Browse for a location 1 Template Effective Date: 05/27/2019 | HOOD05162003010284 Effective Date: 6/30/2019 | HOOD0516200301019 Unrestricted © Siemens Healthcare GmbH, 2019 Atellica® Solution SIEMENS Restoring Data Job Aid Healthineers To view and restore previous database backups: SIEMENS Healthineers AtellicaTM Solution Version 1.13.0.2884004 P-P1-03 1 On the Command bar, select Maintenance System Samples nventory Calibration 2 Qc Patient Orders Worklist Maintenance Setup 1. Schedule Report Log IMT Data Restore Software Updates Copy Configuration 2. Select Data Restore Select Database Backup Current Database Version 1.13.2884.4 Browse 3 Path F.WUSERROOT\Database Backups\ 3. In Select Database Backup, select the required database Reset Date and Time Type Uncompressed size Data Included File Name Monday, March 12, 2018 12:30:00 AM Automatic 2641 ME PCC P-P1-03_AutomaticDataBackup_2018_03_12_00_30_00 zip • Database backup locations can include a drive on the Sunday, March 11, 2018 12:30:00 AM Automatic 4412 MB PCC, P1-060. P-P1-03. C-P1-06. VMM P-P1-03_AutomaticDataBackup_2018_03_11_00_30_00.zip PCC, P1-060. network Saturday, March 10, 2018 12:30:00 AM Automatic 4401 MB P-P1-03. C-P1-06. VMM P-P1-03_AutomaticDataBackup_2018_03_10_00_30_00 zip 4. Select the following criteria of the database restore: Perform Data Restore A. To overwrite all the data in the file, select All in 4 Module (Serial Number) Exclude from Restore O elected 5 Perform Data Restore VPCC QC Statistics Configuration Data and Results Historical Results (Patient and Control) VMM B. To restore data from specific modules, select Selected P-P1-03 P-P1-03 P1-060 then select the module 1-P1-06 5. To restore all data except specific files, select the data 6 Restore groups to exclude in Exclude from Restore ? PEP LabManager 500 Monday, March 12, 2018 12:13 PM 6. Select Restore... > Yes NOTE: A system restore restores the database to the last saved backup. This backup is a NOTE: If the restore fails, consider excluding any data that has snapshot of the system at the date and time it saves. Anything that occurred on the failed. system after that backup cannot be restored. For example, reagent pack inventories will be lower than the on-system volumes which results in expected Reagent Volume Check NOTE: If a database backup does not include data for 1 or more (RVC) errors. modules, those module selections are disabled because nothing requires restoration. 2 Effective Date: 6/30/2019 | HOOD0516200301019 Unrestricted © Siemens Healthcare GmbH, 2019 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2019 Global Siemens Healthineers Headquarters Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84 0 siemens.com/healthineers 3 Effective Date: 6/30/2019 | HOOD0516200301019 Unrestricted © Siemens Healthcare GmbH, 2019
- attelica
- attellica
- atelica