Biograph mMR Cardiac PET Alignment with MRAC Job Aid

This Job Aid provides workflow to align the u-Map to the PET images using the 3D Fusion in the 3D Task Card.

Biograph mMR Cardiac PET Alignment with MRAC µ_Map Registration – Alignment Tool µ-Map might be misaligned to PET Image (e.g. wrong breath-hold position, cardiac exam) 1. Use the 3D Task Card to create an alignment matrix (e.g. µ-Map vs. NAC-PET) 2. Load µ-Map into 3D task card 3. Fuse misaligned volume (e.g. NAC-PET; load via fusion icon 4. Use 3D Fusion Registration to Align volume to µ-Map Story e Essen 18 Storvion Essen 18 MR Star vibe Ess Exam satLAY 2968043 PT StarMibe Esse Cor Viewirig Type Orien. 385 Filming -0 aulo BATTIED 30 Fusion Registration X Registratin Bogroen Registrasco Musik "Default 3D Visual Alignment RetroRecon F BW11:58 Cancel Helo Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 08/21/2020 | HOOD05162003122002 Healthineers Job Aid | Biograph mMR Cardiac Pet Alignment with MRAC Effective Date: 08/21/2020 | HOOD05162003122002 5. Press OK to save alignment matrix a. A series, 3D Application Data will be created MR-PET VEZ [1] 3D Application Data 188 Pers | voto| 1 30 Application Data _F_COMPO 6. Load PET-raw data into Retro Recon 7. Select created alignment matrix in AC tab 8. Select recon parameters and start recon as usual Lung Compartment Adaptive Alignment Matrix Phantom Scan [1] 3D Application C 2 Compartments PET-based ext. FoV Off Patient Recon AC Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 08/21/2020 | HOOD05162003122002 Healthineers Page 2 of 3 Job Aid | Biograph mMR Cardiac Pet Alignment with MRAC Effective Date: 08/21/2020 | HOOD05162003122002 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 a Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 08/21/2020 | HOOD05162003122002 Healthineers Page 3 of 3