Broad Quantification Calibration Procedures

This step by step procedure will guide the user in performing the Sensitivity Calibration and
Volume Sensitivity Calibration which is necessary for the accurate quantification of Broad Quantification studies. The Sensitivity Calibration measures the difference between the system's ability to detect the radioactivity and the true amount of radioactivity using a point source and collimator. The Volume Sensitivity Calibration generates a volume sensitivity factor (VSF) using a phantom with isotope and collimator.

Broad Quantification Calibration Procedures Step by Step Description: This Step by Step procedure will guide the user in performing the Sensitivity Calibration and Volume Sensitivity Calibration which is necessary for the accurate quantification of Broad Quantification studies. The Sensitivity Calibration measures the difference between the system's ability to detect the radioactivity and the true amount of radioactivity using a point source and collimator. The Volume Sensitivity Calibration generates a volume sensitivity factor (VSF) using a phantom with isotope and collimator. What You Will Need: Sensitivity Source (respective clinical isotope to be calibrated) Yielding a count rate of 5-30 kcts/sec per detector Amount of activity needed to achieve this depends on the respective isotope. Here are some examples: Tc99m – 1-5 mCi o In111 – 1-5 mCi o 123 o I 1-5 mCi – *Lu177 – 10-30 mCi o o I 131 1-5 mCi – Peaking and Calibration Source (respective clinical isotope to be calibrated) Filled Uniform Cylinder Phantom (prepared with a known activity concentration reported from a dose calibrator) Yielding a count rate of 5-30 kcts/sec per detector SIEMENS Healthineers www.usa.siemens.com/healthcare HRT# HOOD5162002575206 1 of 6 Performing the Sensitivity Calibration 1. Before starting these procedures, make sure the following are complete: a. The system is homed b. The system is tuned and peaked with appropriate sources. c. Appropriate collimators are installed. 2. Mount the calibration source a. Insert calibration source into the end of the integrated source holder b. Position activity in the vial over the center of the detectors c. Make sure the count rate is between 5 and 30 kcts/sec per detector. 3. On the syngo MI Acquisition Workplace, register or select the patient and launch the Sensitivity Calibration workflow that is located in the Calibration category. 4. Select the Camera Preset a. Camera Parameters Tab b. Select the appropriate camera preset from the pulldown menu. Select the isotope being used for the calibration 5. Enter the Dose Information. a. Series Information Tab Dose numeric value in µCi, mCi, MBq or KBq Measured time and date Summary Series Information Camera Parameters Stop Conditi Series: Technologist Reading Physician Referring Physician: Organ Isotope: Label 67 Gallium Date: Time Dose: 0 MBq 8/ 7/2016 12:00:00 AM Date Coll Preset |Source D SIEMENS Healthineers www.usa.siemens.com/healthcare HRT# HOOD5162002575206 2 of 6 6 . Acquire the Sensitivity Calibration a. Press Prepare Acquisition System initiates motion to position the detectors with a 10 cm source to collimator distance. b. Press Start Acquires an uncorrected image for both detectors simultaneously. Proceeds to Sensitivity Calculation activity. Camera Parameters Camera Parameters Supine Supine 1024 X 1024 1024 X 1024 1.0 10 SC-Tc99m-NMG SC-Tc99m-NMG Both Detectors Done Both Detectors Done Prepare Acquisition Prepare Acquisition 7. Sensitivity Calculation Activity a. Processes the Calibration data. Review the following: Patient Collimator Isotope Presets Crystal Thickness C. Photopeak menu -> Photopeak Measured sensitivity results are displayed If results acceptable -> Accept These Calibrations and Make Them Current If results not acceptable -> Reject These Calibrations and Use Existing Set Sensitivity Calibration Piove 20170112_Bayf_SereCait Collimator Low EnergyHigh Rescion Photopeak Wien Technetum 140 ha Detector 1 Senotty. Nominal OPSMBA Sonotvy- Nomnat CASADA Senotvity- Measured 03.4 CPSMDG CPSABA Devisbios hors Nominal Deltion from Nominal Reject These Cartasces and Use Exsing Sor Acceck These Canteinoes and Make Them Curves SIEMENS Healthineers www.usa.siemens.com/healthcare HRT# HOOD5162002575206 3 of 6 d. Click Complete Image data and Sensitivity Calculation save screen are saved to the patient database e. Remove calibration source f. Retract integrated source holder. Performing the Volume Sensitivity Calibration 1. Prior to acquiring the Broad Quant Volume Sensitivity Calibration, make sure the following have been completed. a. The system is homed. b. Sensitivity calibration has been performed with a point source of the relevant clinical isotope. c. Appropriate collimators are installed d. A uniform cylinder phantom has been prepared with a known activity concentration, as reported by a dose calibrator. e. The phantom is placed on the patient bed and in the field-of-view. f. Ensure that you have a count rate between 5 and 30 kcts/sec per detector. 2. On the syngo MI Acquisition Workplace, register or select the patient and launch the Volume Sensitivity Calibration workflow that is located in the Calibration category. 3. Enter the Dose Information a Select the Series Information tab b. Enter the following: Dose numeric value and unit ( Ci, mCi, MBq, KBq) • Date and time the isotope was measured • Volume Solution: Volume amount of the solution in the phantom. • Dose Calibrator ID: Select Dose Calibrator used to assay dose. • If applicable, enter the Residual Dose information and Dose Calibrator ID. • Summary Series Information |Camera Parameters Stop Conditions Series: Technologist Reading Physician: Referring Physician: Organ Enable Quantitative Acquisition Radiopharmaceutical Isotope: 67 Gallium Initial Assay: Date: Time: Dose: 0 MBq 8/ 4/2016 12:00:00 AM Volume of Solution: 6500 m Dose Calibrator ID Residual Dose: Dose: 0 MBq 8/ 4/2016 12:00:00 AM Dose Calibrator ID Corrected Dose: Dose: MBq SIEMENS Healthineers www.usa.siemens.com/healthcare HRT# HOOD5162002575206 4 of 6 4. Set the Acquisition Parameters a. Stop Conditions tab Stop Condition (counts - default is 100 kcts/view) • Orbit (Circular or Noncircular) • Summary Series Information Camera Parameters Stop Conditions Camera Parameters, Matrix Size 128 X 128 Zoom 1.23 Camera Preset Tc99m-NMG Detectors: Both Detectors Apply Uniformity Correction Orientation Body Position: Supine 5. Acquire the Broad Quant Volume Sensitivity Calibration a. Press Prepare Acquisition The system enables the Start button if all data and parameters pass system validation. • b. Press Start The SPECT acquisition begins. • Upon completion, the workflow proceeds to the CT Acquisition activity. • c. CT Acquisition • Click Prepare button • Perform the CT Acquisition with the parameters set by the workflow. Camera Parameters Camera Parameters Supine Supine 1024 X 1024 1024 X 1024 1.0 1.0 SC-Tc99m-NMG Done SC-Tc99m-NMG Both Detectors Both Detectors Done Prepare Acquisition Prepare Acquisition 6. Volume Sensitivity Reconstruction a. Volume Sensitivity Reconstruction consists of 2 xRecon activities 1 st xRecon activity - used for creating a SPECT image for registration. o o Image Registration activity used to review registration and readjust alignment , if needed. 2 nd xRecon activity – used for final image reconstruction. o 7. Volume Sensitivity Processing Activity SIEMENS Healthineers www.usa.siemens.com/healthcare HRT# HOOD5162002575206 5 of 6 a. Pre-positioned spherical VOI is initially shown. b. Center and resize VOI if necessary. Double-click within the VOI boundary to enable editing. • Single click outside to complete the edits. • c. Avoid phantom boundaries Information Displayed: Collimator, Isotope, Camera Preset • VOI statistics • Calculated Volume Sensitivity Factor • True Activity Concentration • 20160623_Bay8_VSF_Ga67, 16.06.23-19:09:42-DST-1.3.12.2.1107.5.6.1.1008 Elle View Templete Workflow Actvides Help VODAF *DASS Display CT Regrastion Volume Sensitivity Processing Colimenor: Medium-Energy, Low-Penetration Inotopo: 87 Gatium Preset SC-Ga67-N/G Coronal 100 Frustration RICOH Smitty Pitcessing VOI Results: igeble Estoy Parameter Value Statistics Max 38347 58 Boml Min 22724.79 Boml 29583 87 Bgirl True Activity Concentration 25.24 kohl Std Dev. 1640.92 Vol 1628.98 ml Volume Sensitivity Factor 1.17 x Size 158.45 mm Y Size 161:54 mm Z Size 121:54 mm Camplets Suspend Setup Reject These Calbrations and Use Existing Set Accept These Calibrations and Make Them Cument Avg: 0 00 Balml (28 .- 30) Thick: 1 95 mm Accept These Calibrations and Make Them Current button • -OR- Reject These Calibrations and Use Existing Set button • Reject These Calibrations and Use Existing Set Accept These Calibrations and Make Them Current *xSPECT Quant 177Lu is not commercially available in some countries, including the US. 177Lu is not currently recognized by the US FDA as being safe and effective, and Siemens does not make any claims regarding its use. Due to regulatory reasons, its future availability cannot be guaranteed. Please contact your local Siemens organization for further details. SIEMENS Healthineers www.usa.siemens.com/healthcare HRT# HOOD5162002575206 6 of 6

  • Broad Quantification Calibration Procedures