Cios Family - Cleaning and Disinfecting
The job aid details the cleaning and disinfecting steps for Cios Alpha (VA20), Connect, Fit, Fusion, Select and Select FD.
Cleaning and Disinfection Job Aid Cios Alpha (VA20), Cios Connect, Cios Fit, Cios Fusion, Cios Select, Cios Select FD Before cleaning or disinfecting a Cios system, ensure that the system is disconnected from the power supply and switched off. Cleaning the System Parts • Clean the system parts with a damp cloth. • For moistening, use water or a lukewarm diluted aqueous solution, consisting of water and a household cleaning agent. • Cleaning agents and disinfectants containing chlorine derivatives can discolor cables but will not negatively impact system functions. • Never immerse system parts (except the footswitch) in liquid or autoclave system parts. Cleaning the Screen Surfaces/TFT Displays The screen surfaces can be cleaned at any time. • Clean the monitor screen with a cotton cloth dampened with water. • Immediately dry off the monitor screen with a soft cotton cloth. • Wipe off contrast agent spots as soon as possible. Disinfection • For the disinfection of surfaces, we recommend liquid solutions of common surface disinfectants based on aldehyde and/or amphoteric surfactants (e.g. Tensodur 103, Kohrsolin, Cidex). • Certain substituted phenol-based or chlorine-splitting disinfectants can corrode materials and are therefore not recommended. • Only use the recommended cleaning agents and disinfectants. • Do not allow any cleaning fluids to enter the openings of the system (e.g. air openings, gap between covers). • Observe the operator manual of the disinfectant. • Some substances contained in disinfectants are known to be hazardous to health. The concentration of such substances in the air must not exceed the legally defined limit. We recommend that you follow the manufacturers’ instruction for use for these products (active ingredient class). Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 03/23/2021 | HOOD05162003176844 Healthineers Job Aid | Cleaning and Disinfection Effective Date: 03/23/2021 | HOOD05162003176844 Recommended Active Ingredient Classes The following active ingredient classes may be used to clean and disinfect your Cios system: • Aldehydes, alkylamines, quaternary compounds, guanidine derivatives, peroxide compounds, organic acids, chlorine derivatives, alcohol, benzene, common household dishwashing liquids, detergents. Protection Classes and Protection Measures System Component Protection Class Protection Measures X-ray tube assembly (single tank) IPX3 waterproof Protect with sterile cover Flat detector IPX0 no protection Protect with sterile cover Image intensifier IPX0 no protection Protect with sterile cover C-arm* IPX0 no protection Protect with sterile cover IPX3 waterproof Protect with sterile cover Standard footswitch IPX8 protects against immersion Separate from chassis before beyond 1m cleaning Hand switches IPX3 dust protected Separate from chassis before Protected against splashing of cleaning water Wireless footswitch** IPX8 protects against immersion Separate from chassis before beyond 1m cleaning * To find out which system fits to which protection class, please refer to the respective operator manual. **Optional Reference: The operator manual of the relevant Siemens Cios system Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 03/23/2021 | HOOD05162003176844 Healthineers Page 2 of 3 Job Aid | Cleaning and Disinfection Effective Date: 03/23/2021 | HOOD05162003176844 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautio ns. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples o nly. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 03/23/2021 | HOOD05162003176844 Healthineers Page 3 of 3
- cios
- cleaning
- disinffect