Cios Spin Retina 3D Workflow Video
This video demonstrates how to perform a 3D scan with the 3D option at the Cios Spin system.
Welcome to the CR spin Retina 3D workflow video. In this video you will learn how to perform image acquisition with the optional Retina 3D feature of the CS Pin system. As a prerequisite for a 3D scan, you have to register a patient. A fluoro image should be done to check if the see Arm has been positioned correctly in relation to the patient and if the volume of interest is in the isocenter. Set the C arm to zero degrees. Position, angular and zero degrees position orbital or an adequate AP position to be able to see the volume of interest. Perform the horizontal and vertical adjustment with the detector laser light localizer and the horizontal laser light localizer. Press the laser light localizers button on the keypad of the detector housing to turn the laser lights on. The crosshairs allow the adjustment of the system position without releasing radiation. Lock all breaks by pressing the respective buttons on the keypad of the detector housing. Select 3D scan on the control panel. Please note that you can select the 3D scan button only after registering a patient. In the 3D patient scan dialog, you can adjust the examination settings under configure scan. Move this slider to change the scan quality if necessary. Please note if you work with the Navi link 3D option, you can select the correct navigation unit here. When you have configured the 3D scan, press the patient system relation button on the workflow bar. Select the correct patient orientation depending on the patient's position on the table in the operating room. For example, prone when the patient is lying on his or her belly for an examination of his or her spine. Set the C arm position in relation to the patient by selecting one of the respective icons. For example, select this icon. If the see arm is positioned on the right side of the patient. When the patient position in relation to the device has been defined, you can perform the collision check. This check has to be performed prior to a 3D scan to ensure no items are in this scan path. Press the collision check button on the workflow bar to open the collision Check page. Move the see arm to the collision check start position which corresponds to the system and position of the actual 3D scan. When the C arm has reached the collision check start position for a full scan. This position is automatically accepted as the end position of the 3D scan and will be indicated by a check mark. Move the see arm to the collision check and position which corresponds to the start position of the 3D scan. You can perform the collision check either manually by hand or with motorized movements. Please stay next to the patient during the collision check and keep the movement button pressed. Stop the collision check if the patient or any medical devices are in the scan path after the successful collision check the system is ready for the 3D scan and the scan page opens automatically. It is necessary to start the 3D scan within 180 seconds after the collision. Check. The remaining time is displayed on the scan page. Start the 3D scan by pressing the radiation release pedal on the footswitch. The current assignment configuration of the Footswitch pedal is displayed in the control area of the Left Monitor and on the control panel. During a radiation, the radiation display on top of the monitor trolley lights up. Keep the radiation switch pressed until the scan progress bar on the control panel disappears and the reconstruction in progress window opens. When the reconstruction has been completed, the images will be displayed on the left monitor. The 3D patient scan dialog will be closed and the 3D image processing mode will be activated on the control panel.
r, Spine 123456 Orb. 00 Orb.OO Orb. 050 Orb. 950 Orb. 0950 Orb. -590 Orb. -950 orientation Ortho Sphe Store Start Select scan quality Select patient orientation Move C-arm manually to: Collision check Start position Press radiation now to start 3D scan Press radiation switch now to start 3D scan Ang.00 position Relation Set the C-arm position Select system position Select navigation Remaining time: 110 s 80 Lock 23 Patient orientation Select patient orientation patient orientation in relation to the patient pGym2 Remaining time: 177s 17 No Navigation Reconstruction Patient orientation Select patient orientation Scan quality posigation System position Select system position Set the C-arm position High quality hh:mm:ss mAs Store Start 3D Scan progress Orbital 970 7! Extra power: Obese No Navigation 3D Orbital -990 BrainLAB '630s 400 Images Reset 30s X 100 Images 60s 4 100 Images Laser Supine Prone 3D Angular 00 Scan timet timer tirner timner timnet
- ciosspin
- c-arm