Cleaning System Fluid Drawer on the Atellica® CH Analyzer Job Aid Version 1.25 and Below
This job aid provides instructions on how to perform the as needed maintenance task of cleaning the system fluid drawer for the Atellica® CH Analyzer.
Atellica® CH Analyzer SIEMENS Cleaning System Fluids Drawer Job Aid Healthineers 1. Enter the system into Diagnostic State: A. Navigate to System > Operator Diagnostics > CH Icon B. Select Enter Diagnostics > Yes 2. Wait for analyzer to enter Diagnostic State 3. Under Subsystems, select Inspect and Clean 4. Select Cleaning the System Fluids Drawer > Perform > Yes 5. Wait for the analyzer to prepare for maintenance 6. Locate and open the system fluid drawer 7. Tighten the cap on the system fluid containers Doc 8. Remove the system fluid containers 9. Carefully wipe down the system fluid drawer with lint-free tissues or gauze saturated with special reagent water 10. Replace the system fluid containers Note: Do not touch the cannulas in the keyed areas at the bottom of the system fluids drawer. The cannulas can puncture skin. *This procedure is continued on the next slide. 1 Template Effective Date: 05/27/2019 | HOOD05162003010284 Effective Date: 24-SEP-2020 | HOOD05162003126734 Unrestricted © Siemens Healthcare GmbH, 2020 Atellica® CH Analyzer SIEMENS Cleaning System Fluids Drawer Job Aid Healthineers 11. To prevent a vacuum from forming inside the container that could cause the analyzer to stop processing, vent the container. To vent the container, squeeze and twist the cap counterclockwise to pass the first designated stop. Then no longer squeeze, but twist the cap until it reaches the next designated stop 12. Close the drawer 13. Select Continue > Close 14. Select Exit Diagnostics > Yes 15. Mark the procedure as completed in the Maintenance > Schedule screen 2 Template Effective Date: 05/27/2019 | HOOD05162003010284 Effective Date: 24-SEP-2020 | HOOD05162003126734 Unrestricted © Siemens Healthcare GmbH, 2020 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2020 Global Siemens Healthineers Headquarters Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84 0 siemens.com/healthineers 3 Template Effective Date: 05/27/2019 | HOOD05162003010284 Effective Date: 24-SEP-2020 | HOOD05162003126734 Unrestricted © Siemens Healthcare GmbH, 2020
- attelica
- attellica
- atelica