Cleaning the Exterior of the Aspirate Probes on the Atellica® IM Analyzer Video Version 1.25 and Below

This video provides instructions on how to perform the monthly maintenance task of cleaning the exterior of the aspirate probes on the Atellica® IM Analyzer.

exterior of the aspirate probes monthly to begin, but the analyzer and stop mode by selecting the system button on the command bar. Next, select the status tab. Then the immuno assay module graphic. On the immuno assay status screen, select the stop or pause icon. Once the analyzer is stopped, open the front cover by first selecting the icon to unlock the front cover. Wait for the lock graphic to indicate the appropriate state, then open the front cover. The system has four aspirate probes, aspirate probe, one aspirate probe, two aspirate probe, three, an aspirate probe 4 all four probes are identical and the procedure is the same to clean each of them, carefully lift the probe arms to the highest position. Note when working with the aspirate, probes do not remove the probe. Also, be careful not to touch the probe tip and do not bend the probe. Clean the visible portion of each probe using a gentle downward motion with lint free tissues or gauze moistened with IM cleaner. And wipe each probe using a gentle downward motion with lint free tissues or gauze saturated with special reagent water. Repeat rinsing three to five times with clean lint, free tissues, or gauze. Close the front cover. To complete the procedure, start the analyzer by selecting the resume icon on the Immunoassay System status screen. Then Mark the cleaning the exterior of the asparate probes activity is completed by selecting the maintenance button on the command bar. Then the schedule tab. On the schedule screen, select the cleaning the exterior of the aspart probes activity. Then the mark as completed button on the right hand side of the screen.

SIEMENS Build 1.7.0 SIEMENS Build 1.7.o SIEMENS Build 1.2.0 DL-P1-07 System Samples Inventory Calibration Patient Orders Worklist Maintenance Status Setup Schedule Seatup Status Logs Operator Diagnostics Log Logs Diagnostics About Copy Configuration Failter Module: All Frequency All Time: Next 24 hours Performed By: All Module: All Frequency: All Time: Next 24 hours Performed By: All System VMM p -p 1-04 p -PI .04 p.p1-04 DL-P1-07 pl.02 Ferforrr Perform. Module Activity Schedule Fri-14 Sat-15 Mon-17 Tue-18 Wed-19 Thu-20 Tue-18 Mark as Completed Standby Ready Stopped Stopped Stopped p p_lA-O I Cleaning the Exterior of Reagent Probes Cleaning the Exterior of the Aspirate Probes Cleaning the Exterior of Rheagenirate Probes 4/13/2017 4/13/2017 1:00 Not Perform___ Capabilities Alert Category Time Cleaning the Water Container Procedure Details Wash Ring Search Text Enter the activity name. p p _IA-O I Inspecting and Emptying the 1M Water Trap 4/17/2017 4/13/2017 4/13/2017 1 oo Luminometer EditNiew__ EditNiew___ Cle ,ning the Exterior of the Scheduled by Operator Schedul d by Operator Inspecting and Emptying the 1M Dryer Trap Sample Probe Scheduled Automatically Schedül d uto atica 4/18/2017 3:00 UPS Scheduled by Service DL-P1-07 Incubation Ring Incubation Ring Thermal DL-P1-07 Connected to PCC Failed •.-Reagent Compaäment Tnermal Reagent Compartment Thermal 4/18/2017 0:30 Overdue Luminometer Thermal Inprocess (b Below Deck Heating upper Deck Heating Upper Deck Heating Completed p p _IA-O I 1M Daikly Maintenance Incubation Ring Diagnostics Not Perform___ Not Performed_ pp_lA-Ol 1M Weekly Maintenance 1M Daikly Maintenance 4/20/2017 5:00 Tip Loader With Comments Patient Sample Processing •./ Patient Sample Processing 5/1/2017 TOO 5/17/2017 10:05 5/18/2017 0:30 Below Deck Heating Front Cover Back Cover pp_lA-Ol Reagent Probe 1 5/17/2017 10:05 4/17/2017 3:30 Front Cover Back Cover Front Cover Add Activity _ Set Warning___ Set Warning.__ PEP LabManager Thursday, June 08, 2017 1 AM Thursday, June 08, 2017 AM