PEPconnect

Cleaning the Atellica® Sample Handler Connect System Exterior Job Aid Version 1.25 and Below

This job aid provides instructions on how to perform the as needed maintenance task of cleaning the Atellica® Sample Handler Connect system exterior.

Atellica® Sample Handler Connect SIEMENS Clean the System Exterior Job Aid Healthineers 1. Clean the exterior surfaces with lint-free tissues saturated with 10% bleach solution or gauze saturated with a 10% bleach solution 2. Wipe the exterior surfaces with lint-free tissues saturated with water or gauze saturated with water 3. Dry the exterior surfaces thoroughly with clean lint-free tissues or gauze HOOD05162002822811 Effective Date: 3/12/2018 Page 1