Performing Maintenance - Cleaning Job Aid
This job aid provides instructions on how to perform Maintenance on your CLINITEK Advantus® Analyzer, including various cleaning and disinfecting activities.
Maintenance - Cleaning Clean the Exterior and Touch Screen Remove and Clean the Push Bar 1. Turn the analyzer power OFF 1. Turn the analyzer Off 2. Clean the exterior of the analyzer using a damp 2. Tilting the bar slightly upwards, pull the bar cloth and a mild detergent straight out 3. Clean the touch screen, with a soft nonabrasive 3. Clean the push bar with warm water and a mild cloth dampened with a mild glass cleaner detergent • Do not spray cleaner directly on the screen, or use bleach to clean the display 4. Rinse thoroughly and dry with a paper towel or soft cloth 5. To reinstall the push bar refer to the Maintenance: Reinstalling tab in this document SIEMENS GPE04001.005 Effective Date: 09/12/2014 Maintenance - Cleaning Remove and Clean the Fixed Platform 1. Turn the analyzer Off 2. Remove the waste bin container and discard the used test strips into an appropriate container 3. Remove the fixed platform by pull the entire assembly towards you 4. Clean the platform with warm water and a mild detergent • Rinse thoroughly and dry with a paper towel or soft cloth • Allow the calibration bars to air dry 5. Inspect the calibration bars for scratches, marks, or discoloration • Discard and replace the platform with a new one if the bars can not be cleaned 6. To reinstall the platform refer to the Maintenance: Reinstalling tab in this document GPE04001.005 Effective Date: 09/12/2014 Maintenance - Cleaning Remove and Clean the Holddown Plate Remove and Clean the Moving Table 1. Turn the analyzer Off 1. Turn the analyzer Off 2. Press upwards on the tab at the back of the plate 2. Pull the moving table towards you 3. Pull the other end from its retaining hole 3. Clean the moving table with warm water and a 4. Clean the holddown plate with warm water and a mild mild detergent detergent • Rinse thoroughly and dry with a paper towel or • Wipe with a damp cloth soft cloth If the holddown plate is extremely dirty, soak in warm • Use care when drying around the pins • water and a mild detergent to loosen the dried 6. To reinstall the moving table refer to the residue Maintenance: Reinstalling tab in this document • Rinse and dry with a paper towel or soft cloth 5. Use a cotton-tipped swab, wetted with plain water 6. To reinstall the holddown plate refer to the Maintenance: Reinstalling tab in this document
- Maintenance
- Cleaning
- Test Table insert
- routine
- periodic
- disinfecting
- calibration bar