CLINITEK Atlas® Automated Urine Chemistry Analyzer Calibration Online Training
Identify the conditions under which the system will require a calibration. Prepare the calibration materials. Perform a calibration. Review and print the calibration status. List the steps to resolve common calibration errors. This clinical laboratory training qualifies for continuing education units (CEU).
Magnetic Susceptibility increases linearly at 3T. Some of the challenges are increased in-plane spatial distortions and increased through-plane excitation errors. Magnetic susceptibility has a larger affect on GRE sequences due to the use of gradient refocusing pulses instead of RF refocusing pulses and higher sensitivity to T2 star effects. Welcome to the CLINITEK Atlas® Automated Urine Chemistry Analyzer Calibration Online Training course. This course includes information regarding when to calibrate, how to run calibrators and how to review the results. Select Next to continue. Identify and resolve common calibration errors Identify the conditions under which the system will require a calibration List the steps necessary to perform a calibration Locate and review the status of calibration results State the materials required for calibration Upon successful completion of this course you will be able to: Select Next to continue. If a 3D sequence is acquired with an isotropic voxel, the data may be reformatted using Multi-planar reconstruction. Multi-planar reconstruction is a post processing reformatting of a 3D data set into a different imaging plane than what the sequence was acquired in. In this example, it is showing a 3D Axial DESS knee that has been loaded into the 3D MPR card and reformatted into the Sagittal plane. Congratulations. You have completed the CLINITEK Atlas Automated Urine Chemistry Analyzer Calibration Online Training course. Listed below are the key points that have been presented. Take time to review the material before you proceed to the final quiz. Identify the conditions under which the system will require calibration Calibration on the CLINITEK Atlas analyzer is required when: A reagent roll with a different lot number is loaded. A reagent roll with the same lot is loaded but the system has not been calibrated within the last 24 hours. The system displays error messages requiring you to calibrate the system. The system temperature changes by more than 5° C. State the materials required to calibrate Sample tubes CLINITEK Atlas Calibration Kit Tube of bleach (5.25% sodium hypochlorite concentration) Sample rack List the steps necessary to perform a calibration Bring the calibrators to room temperature Select Analyze Select Calibrate Verify Lot numbers Load the calibrators Select Continue Locate and review the status of calibration results Review the Calibration Successful screen when a calibration passes. Print a calibration report by accessing the Main Menu and selecting Print QA Report. Identify and resolve common calibration errors Common calibration errors include: Insufficient Calibrator volume: Add additional calibrator to the tubes. Dry Calibration failure: Recalibrate, replace the reagent roll, clean the readhead or replace the lamp. Wet Calibration failure: Confirm that the calibrators were handled properly and have not expired, verify pipette performance, check the reagent roll expiration, or clean the readhead. Prime Pump: Select Prime Pump on the error screen or access the Main Menu and select 5: Prime Pump. Pad Dispense error: Verify pipette performance or replace reagents. Select Next to continue. This is Prof. Roentgen’s lab in the Physics building at the University of Wurzburg in Wurzburg, Germany. There is little to denote what this room is, except for a bust of Roentgen in the hallway. The CLINITEK Atlas analyzer calibration is required: Any time a reagent roll with a different lot number is loaded When a reagent roll with the same lot is loaded but the system has not been calibrated within the last 24 hours If the system displays error messages requiring a calibration of the system If the system temperature changes by more than 5° C After hardware components are replaced and it is required by the procedure Select Next to continue. Materials required to calibrate the system include: Sample tubes CLINITEK Atlas Calibration Kit Tube of bleach (5.25% sodium hypochlorite concentration) Sample rack Preparing to Calibrate Learn about calibration preparation. Slide NumberText BlocksCalloutsAudio ScriptImage File1 Calibration Kit Calibrators must be run at room temperature. Store the kit at 2 - 8°C. Check the thermal integrity card inside the box of calibrators to ensure the calibrators have not been exposed to excessive heat. Select Next to continue. Note: If audio does not automatically start, select the play arrow in the top left to begin.Remove the calibration kit from the refrigerator and allow the calibrators to reach room temperature. Store the kit at 2 - 8 degrees Celsius when not in use. Check the thermal integrity card inside the box of calibrators to ensure the calibrators have not been exposed to excessive heat.2 Prepare Bleach Solution Pour at least 2 mL of 5.25% sodium hypochlorite into a sample tube. If the concentration of sodium hypochlorite is higher than 5.25%, dilute the bleach with distilled water. For example, if the bleach is 6% sodium hypochlorite, add 0.75 mL of water to 5 mL of the 6% bleach and gently swirl to mix. Select Next to continue. Pour at least 2 mL of 5.25% sodium hypochlorite into a sample tube. If the concentration of sodium hypochlorite is higher than 5.25%, dilute the bleach with distilled water. For example, if the bleach is 6% sodium hypochlorite, add 0.75 mL of water to 5 mL of the 6% bleach and swirl to mix gently.3 Prepare Sample Tubes Gently mix the calibrator bottles Pour at least 3 mL of calibration solutions 1, 2, and 3 into sample tubes labeled CAL 1, 2, 3. Pour at least 5 mL of calibration solution 4 into a sample tube labeled CAL 4. Allow the calibrators to come to room temperature before placing on the system. Select Next to continue. Gently mix the calibrator bottles. Pour at least 3 mL of the calibration solutions 1,2 and 3 into sample tubes marked 1, 2, 3. Pour at least 5 mL of calibration solution 4 into a sample tube labeled 4. Allow the calibrators to come to room temperature before placing on the system.4 Load Tubes Place the sample tubes in the rack in the following order: Bleach - pos. 1 Cal 4 - pos. 2 Cal 1 - pos. 3 Cal 2 - pos. 4 Cal 3 - pos. 5 Place this rack onto the right side of the Sample Handler, with the open side facing towards the back When complete, select the X in the upper-right corner to close the window and continue. Place the sample tubes into the rack in the following order: Bleach in position 1, then Cal 4, Cal 1, Cal 2, and Cal 3, as depicted on the screen of the Calibration Procedure A calibration can be initiated by: Selecting the Calibrate softkey Responding to the instrument prompt when a new lot of reagent roll is loaded Steps during calibration: The SG Well is cleaned with the 5.25% bleach A dry calibration is performed which checks reagents and records baseline readings A clarity calibration is performed using calibrator 4 in the SG Well If clarity passes, calibrators 1-3 are aspirated in duplicate Performing a Calibration Learn about performing a calibration. Instructions:If media does not automatically start, select the play arrow to begin.Flash File:/content/generator/Course_90007042/sim_ATL_Cal_HTML5_800x600_9/sim_ATL_Cal_HTML5_800x600_9.swfHTML5 File:/content/generator/Course_90007042/sim_ATL_Cal_HTML5_800x600_9/index.htmlPDF File: Lys fra lyddioderne følger lyslederen og reflekteres herefter fra enten kalibreringsfelt, testfelt eller kassette på spejlet ved specifikke bølgelængder (470 nm, 525 nm, 565 nm, 625 nm, 660 nm og 845 nm). Intensiteten af det reflekterede lys er afhængig af graden af farveændring i testfeltet og er direkte relateret til koncentrationen af den komponent der testes for. Derefter ledes lyset gennem en åbning på linsen ogfokuseres på detektoren. Den detekterede lysintensitet omdannes til elektriske impulser, som omregnes til kliniske resultater af instrumentes mikroprocessor. You have just learned about performing a calibration. Select the link below to practice what you have learned. Performing a Calibration Practice performing a calibration. Instructions:If media does not automatically start, select the play arrow to begin.Flash File:/content/generator/Course_90007042/sim_ATL_CalYT_HTML5_800x600__9/sim_ATL_CalYT_HTML5_800x600__9.swfHTML5 File:/content/generator/Course_90007042/sim_ATL_CalYT_HTML5_800x600__9/index.htmlPDF File: Common calibration errors include: Low Calibrator Volume Dry Calibration Failure Wet Calibration Failure Prime Pump Pad Dispense Error Handling Calibration Errors Learn about handling calibration errors. Tab TitleTextLow Calibrator Volume If there is insufficient Calibrator volume this message will appear on the screen. The calibration will have to be restarted after additional calibrator is added to the tube or tubes. Prime Pump When a new reagent roll is loaded, rinse solution will need to be prepared. Carefully remove the canula when transferring it to avoid introducing an air bubble. If there are bubbles in the rinse solution line, this error may occur. Select Prime Pump on the error screen. Alternately, access the Main Menu and press 5. Prime Pump Pad Dispense Error When this error occurs it could indicate a problem with the pipette or reagents. Verify pipette performance by completing the Checking for Leakage maintenance procedure. If the pipette is not performing correctly, it may need to be replaced. Refer to the CLINITEK Atlas analyzer operating guide for more information. If the pipette is performing properly, another step to resolve this issue is to replace the reagent roll. Dry Calibration Failure This message is generated when the results from the dry calibration of a dry reagent strip are out of range. It can be related to the reagent pack quality, the readhead or the lamp. Recalibrate the system. If the problem persists, replace the reagent roll with a fresh, sealed roll pak and attempt calibration again. Refer to the CLINITEK Atlas analyzer operating manual to clean the readhead or replace the lamp as needed. Wet Calibration Failure This message can appear if: The calibrators were not at room temperature when analyzed on the system The calibrator tubes were loaded into the wrong rack positions The incorrect calibrator was poured into the labeled tube The calibrators have expired The reagent roll has expired There is an issue with the pipette To resolve a Wet Calibration Failure: Confirm the calibrators are at room temperature. Check the expiration date of the reagent roll and calibrators. Verify the tubes are in the correct positions and poured into the proper tubes. Verify pipette performance by completing the Checking for Leakage maintenance procedure. If the pipette is not performing correctly, it may need to be replaced. Check the readhead and clean if necessary. When complete, select the X in the upper-right corner to close the window and continue. Reviewing the status of the calibration results: When a calibration passes, the Calibration Successful screen is displayed. If a calibration fails, the Problem Reported screen is displayed. A calibration report can be printed from the Main Menu by selecting QA Report. Review Results Learn about reviewing calibration results. Tab TitleTextCalibration Successful Screen If a calibration has passed, the Calibration Successful screen is displayed briefly. The display will then return to the Ready for Testing screen. Problem Reported Screen If a calibration failed, the Problem Reported screen is displayed briefly. The display will then list the specific problem along with any action needed to resolve the issue. When complete, select the X in the upper-right corner to close the window and continue. Printing a Calibration Report Learn about printing a calibration report. To print a calibration report: Access the Main Menu Select Print QA Report The calibration status of Passed or Failed will be displayed on the report. When complete, select the X in the upper-right corner to close the window and continue.
- clini
- calibrator
- clinical laboratory training