CLINITEK Novus® Analyzer Cleaning the SG Well Video
This video provides instructions on how to clean the SG well on the CLINITEK Novus® Analyzer, including how to prepare the sample rack.
The SG well cleaning message appears in the status log when it's been 24 hours since the last time. The SG well was cleaned, the status of the clinic Tech novice is not ready. Patient samples and controls cannot be run until the SG well is cleaned. From the Home Screen, select system to navigate to the procedure. From the system menu, select Clean SG well. Prepare the rack as stated in the instructions here. When preparing the bleach, the concentration should not exceed 5.25% sodium hypochlorite. Place the rack on the rack handler and then select start. The cleaning cycle takes about 3 minutes to complete. You won't be able to navigate to other menus until the cycle is complete. A progress bar appears to indicate the amount of time left. When the cleaning cycle is complete, the software returns to the system menu. From here, select the home button to return to the home screen. The SG well cleaning message is gone and the system is now ready for testing.
SIEMENS Status: Ready CLINITEK Novus Status: Not Ready Status: Processinq Home System > Clean System 51 AM 05/17/2012 12 05/21/2012 12 PM 05/21/2012 12 2B PM 05/21/2012 01 12 2B PM FM Tests Remaining Tes&s Remaining Status Log Tes& Remaining 376 Calibration Diagnostics Diagnostic The cieaning cycle is in process. 1. Place the bleach in position 1. Case & Cesse&te Cassette 10 Test Page SG well cleaning is required. Z. Place the rack. on the rack handler . Cassette Lat Casse& Lot 1022000 Casse&E Lot Onboard Stabilitf 3. Select Start. Corsiyo' Control 13 Days 10 nays System Management 1/1 Bl«h Next Samnple 0-00172 Next Sample use 5.25% in use 525% in The cleaning cycle takes Next STAT s-00001 STAT position 1 _ about three minutes to Clean SG Well SG well SG Well About the System complete. select Clean SG well Prime Pump Prime Pump — Cancel Load & Unload 100% Select the Home button Select System Start Results Se tup Start System