Cleaning the Test Table and Insert Video
This video provides instructions on how to clean the Test Table and Insert on your CLINITEK Status®+ Analyzer as part of maintenance activities. Instructions include how to remove both the test table and the insert prior to cleaning.
To keep the analyzer operating properly, the test table, an insert must be kept clean. For routine cleaning, remove the insert and clean thoroughly. Rinse both sides of the insert under running water, then dry and replace the insert. Periodically or is needed. Clean the test table. Begin by pulling the test table slowly out of the analyzer. Lift the insert from the test table and if needed, drain the drip tray. What a cotton tip. Stick with water and clean the test table. The white calibration bar will be cleaned in a separate procedure. Use a soft cloth or lint free tissue to thoroughly dry the table. Hold the test table at the opposite end from the calibration bar an with the white bar facing upward. Reinsert the table back into the analyzer, pushed the test table firmly and slowly just halfway into the analyzer. Then replaced the insert.
Cleaning the Test Table and Insert