Performing an hCG Quality Control Test Job Aid
This job aid provides instructions on how to perform a Quality Control test for hCG on your CLINITEK Status®+ Analyzer.
CLINITEST® hCG Quality Control Performing Quality Control (Connect System) 1. Verify the test table insert is in position for a cassette test 2. Select QC Test 2. Select QC Cassette Test Required 4. Select Enter lot and expiration date • Enter the lot information using the touchscreen and select Enter to continue 5. Place CLINITEST® hCG cassette into the test table 6. Select START 7. Add control (to marked line, approximately 0.2 mL) into the sample well 8. The test table and cassette will automatically be pulled into the analyzer 9. Review if QC Test displays Pass or Fail 10. When complete, remove and dispose the test cassette per laboratory procedure GPE15004.033 Effective Date: 02/14/2014