Configuration and Use of CT Dose Notification: All Biograph and Symbia Intevo Systems
Dose Notification will appear as a pop up message when a planned scan
range would exceed a user-defined dose value. These parameters can be prescribed for every scan range and are configured in the Scan Protocol Assistant.
Configuration and Use of CT Dose Notification All Biograph and Symbia Intevo Systems Description: Dose Notification will appear as a pop up message when a planned scan range would exceed a user-defined dose value. These parameters can be prescribed for every scan range and are configured in the Scan Protocol Assistant. Configuration 1. Open the Scan Protocol Assistant from the CT Examination Task Card: Options à Configura- tion à Scan Protocol Assistant osert View Setup Image | Options System Help Version License ... File Browser SOMATOM LifeNet Maintenance. Auto Expose Configuration Service 12 Somaris/5 - Configuration Panel X File Ca Report Conversion CaScoring DynEva CT e-Logbook Examination Configuration Configuration Osteo CT Neuro PCT Body PCT Pulmo CT Scan Protoco Assistant Scan Protocol Assistant SIEMENS siemens.com/healthineers Healthineers . Configuration and Use of CT Dose Notification Effective Date: 02/05/2018| Quality HOOD05162002763640 2. Select Change Parameters in the “What do you want to do?” step. 2 ---- ------ 3. In step 2, select the protocols in which you would like to configure a Dose Notification. 2 3 5 Chetate! 4 Thre Riech primocola - Shoulder -- --- proncol a -- Que Siemens Healthineers Services Page 2 of 6 Configuration and Use of CT Dose Notification Effective Date: 02/05/2018| Quality HOOD05162002763640 4. In Step 3 of the configuration, enter the desired Dose Notification values for the CTDIvol and/or DLP for each of the exam series. Scan Protocol Assistant 2 3 5 ..... Change parameters Protocol Topogram Scan Recon Auto tasking Trigger Intervention Contrast Including topogram Range Scan name art time Bolus trigger level CTDIvol TI |X-CARE |Uppe LO .. Dose Notification CTDIvol Dose Notification DLP max KV min kv CARE KV Abdomen / AbdMultiPhase (Adult) Topogram: 35mA, 512mm, AP, Craniocaudal Non Cont. itton 5.15 12.6396 0.0 off 688 8 Arterial itton 5.15 14.1685 0.0 off 140 Venou 5.15 12.6396 0.0 140 off 888 Select all Sequence scans Spiral scans Multiscans Interventional scans Adaptive 4D spiral scans III-Sequence i-Spira -Fluoro CARE Dose type CARE DOS4D FAST Adjust Dose modulation Upper limit max. scan timeal 20.0 s CARE KV On Lower limit max. mAS 25 % Quality ref. mAs Dose Notification CTDIvol Eff. mAs DLP 8 8 Ref. KV 120 kV 120 CARE KV min. KV -180 Dose saving optimized for. max. kV - 140 Dose Parameter Siemens Healthineers Services Page 3 of 6 Configuration and Use of CT Dose Notification Effective Date: 02/05/2018| Quality HOOD05162002763640 UsIng Dose Notification 1. If the user attempts to load a scan range that would result in a dose that exceeds the configured Dose Notification thresholds, a pop up will notify the user. It is still possible to Load the examination. The dialogue box includes an optional field for the user to document a diagnostic reason for the requested exposure (e.g. obese patient). Dose Notification X Arterial Phase: DLP (38.8 mGy"cm) exceeds the notification value (12.0 mGy*cm). Venous Phase: CTDIvol (17.79 mGy) exceeds the notification value (10.00 mGy). Venous Phase: DLP (420.6 mGy*cm) exceeds the notification value (12.0 mGy*cm). Diagnostic reason Load Cancel 2. If Dose Notification values have been configured but the Dose Notification pop up has been disabled in Optionsà Configurationà Examination, the system will display the proposed CTDIvol and DLP in red if they exceed the configured values. 01_AbdMultiPhase (Adult) 10.10.19-15:18:21-DST-Specials PolyTraum 10.10.19-15:18:21-DST-" Total mAs: 184 3D Topogram Eff. mAs CARE DOS04D Non Contrast RT KV 140 CTDIvol (32cm): 19.43 mGy DLP. 459.3 mGy cm Pause Scan time 5.62 5 configured ref value: 10.00 mGy Simulation Deley 25 PreMonitoring 5.0 mm Acq. 128 x 0.8 mm Contrast Monitoring No. of images 45 rterial Phase U Comments Venous Phase Range: Begin End Table: Position Height -863 0 -1084 5 -12880 1250 Craniocaudal Load RACON Recon Routine Scan Recon Auto Tasking No checkup performed so far Select Schup/Check 19-Oct-2010 15:19:32 Siemens Healthineers Services Page 4 of 6 Configuration and Use of CT Dose Notification Effective Date: 02/05/2018| Quality HOOD05162002763640 3. All appearances of the Dose Notification are stored in a retrievable Dose Log found in the Offline File Browser. Patient Applications Edit Insert View Setup Image Options System Help min Examination 17 10.10.19-15:10:21-DST-Specialg En 10.19.15 18:21.081 1 21 H Hospital :10.10.19-15:10:21-DST-Specials ... BOMATOM Definition Biesh 1n 10 19.15-18.21.08. Hospital SOMATOM Dednition Flash Di FileBrowser O X CT 2011A 1.01_POO File Edit CT Data Transfer Help H-8P-CR 8 2 H.\SiteData Name Size | Modified @ cascoring DoseLog_2010-10.txt 547 Byte 10/19/10 15:29:20 +@ CDBum @ ConstancyRepor DoseLog_2010-10.txt - Notepad LO X ContrastProtoco File Edit Format View Help @ DoseLogs Date (dd-mm-yyyy. Time (hh:mm:ss) Patient ID Viewing @ E-Logbook 19-10-2010 15:29:20 Study UID Type Range User name Diagnostic r 10.10.19-15:18:21-DST-1.3.12.2.1107.5.1.4.61999 1.3.12.2.1107.5.1.4.61999.3000001010 19-10-2010 15:29:20 10.10.19-15:18:21-DST-1.3.12.2.1107.5.1.4.61999 1.3.12.2.1107.5.1.4.61999.3000001010 Examination @ ExportedContras @ ExportedScanPr @ ImportContrastP @ @ ImportScanProtc @ InSpace lungcare Filming @ Manuals @ OffLine @ Protocols Referencelmage ScanGuide @ ServicePacks syngoCirculation syngoColon @ syngoOncoCARI 886 3D -B Floppy QL Cn_PW Drive Simulation Venou Range: Begin End Table: Position -1137.53 Height Hold -1359.0 -1111.0 - 125.0 Craniocaudal Load Recon Recon Routine Scan Recon Auto Tasking PNP Device Mounted - Drive(F:): Total Space(152580 MB) 19-Oct-2010 15:52:09 Siemens Healthineers Services Page 5 of 6 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instruc- tions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety infor- mation like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018
- Biograph Symbia Intevo