Creating CT and PET Dose Reports
This job aid will assist the user in creating reports for both PET and CT that reflect the dose that the patient received from the Exam. The Dose Report is created according to the DICOM standard, which summarizes examination data and dose values as a DICOM Structured Report (SR). The Dose Report is created automatically by the system at the end of an examination and can be found in the Patient Browser. This job aid applies to Biograph Vision, mCT (VG7xx) and Horizon scanners.
Creating CT and PET Dose Reports Biograph Vision, mCT and Horizon Description: This job aid will assist the user in creating reports for both PET and CT that reflect the dose that the patient received from the Exam. The Dose Report is created according to the DICOM standard, which summarizes examination data and dose values as a DICOM Structured Report (SR). The Dose Report is created automatically by the system at the end of an examina- tion and can be found in the Patient Browser. This job aid applies to Biograph Vision, mCT (VG7xx) and Horizon scanners. Creating a CT Dose Report 1. Configure the CT Dose Report by enabling it in the following menu: Options à Configuration à Examination à Dose. 2. Make sure the Dose Report checkbox is selected. 3. If you would like the report to be automatically transferred to another system, check the Auto Transfer checkbox. 4. Click Apply to save the changes. Examination Configuration Patient Dose Workflow Topogram Processing Contrast Display Options CARE Dose4D configuration: mAs adaptation to patient size Dose notification CARE Profile v Organ characteristics Child Adult slim Adult obese Brain Average Average Average Exposed range M V Neck Average Average Average Shoulder Average Average Average Dose Report Thorax Average Average Average Activate Dose Report v Auto transfer V Abdomen Average Average Average V Pelvis Average Average Average Additional transfer None Spine Average Average Average Osteo Average Average Average Dose Alert Head/Vascular Head Average Average Average Adul Child Vascular Body Average Average Average V Runoff Average Average Average CTDIvol 1000 m 1000m M Cardio Average Average Average DLF mGy*cm -mGy*cm Respiratory Average Average Average OK Apply Default Settings Cancel Help SIEMENS Healthineers . siemens.com/healthineers Creating CT and PET Dose Reports Effective Date: 07/25/2018 | Quality HOOD05162002915465 Creating a PET Dose Report 1. Configure the PET Dose Report by enabling it in the following menu: Options à Configuration à PET Examination à PET Dose Report 2. Check the Create checkbox to create the report. 3. If you would like the report to be automatically transferred to another system, check the Auto Transfer checkbox. 4. Choose View & Print color. 5. Click Apply to save the changes. PET Examination Configuration X Units Pharmaceuticals | PET Recon Pri Export Series T. Image Statistics PET DoseReport Dose Report Create Auto Transfer View & Print White Background/ Black Background/ Black Text White Text OK Apply Default Settings Cancel Help Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Creating CT and PET Dose Reports Effective Date: 07/25/2018 | Quality HOOD05162002915465 Viewing the Dose Reports 1. In the Patient Browser, you will see the Dose Report and/or the PET Dose Report. 2. Highlight the dose report you would like to view and click the Load Report icon in the toolbar. Patient Browser Patient Applications Transfer Edit View Filter Evaluation Sort Cotions Help [2] Thorax 5.0 B31f [501] Patient Protocol Local Database tost Thorax ThoraxR (502] Dose Report Scheduler drugyuhijk (601] Raw Data sfadwgf 08.10.17-09:32:47-ST test NR 10 15.18 15-30.51 Mon Thorax Thoren Routine (Adun) Date of birth 10/17/1085 Study date: 10/17/2008 Pationt ID 125 Modality SR 10/17/200 -OR- Current Fiter, Off Double click the dose report and the report will load into the Viewing Task Card. Transfer Edit View Image Tools Scroll Evaluation Doe, JOhn Doe, JOhn Doe John 16.09.02-17:05:11-DST-1.3.12.2. Doe JOhn 16.09.02-17:05:11-DST-1.3.12.2. 8/4/1974, M, 42Y 8/4/1974, M. 42Y EXAM 9/2/2016 EXAM 9/2/2016 EXAM 5:05:29 PM EXAM 5:05:29 PM 7 IMA 2001 7 IMA 2002 Viewing AC Viewing PET Dose Report PET Organ Dose Parameter SI units English units Organ mSv Radiopharmaceutical Fluorodeoxyglucose Fluorodeoxyglucose marem Adrenals 2.22 Radioisotope F-18 F-18 Bone surfaces 2.035 Radionuclide Half Life(s) 6586.2 6586.2 Brain 7.03 Filming 185.00 MBq 02-09-2016 Breast 1.628 Injection Date(DD-MM-YYYY) 02-09-2016 Gallbladder wall 2.405 Injection Time(HH:MM:SS) 17:06:35 17:06:35 Stomach wall 2.035 Weight 54.43 Kg 120.00 lbs Small intestine wall 2.22 Height N/A N/A Colon wall 2.405 Blood Glucose N/A N/A Upper large intestine wall 2.22 Effective Dose 7.52 mSv 751.54 mrem Lower large intestine wall 2.59 Tools Image View Tools | Image | View Heart wall 12.395 Kidneys 3.145 Liver 3.885 Lungs Muscles Oesophagus Pancreas Red marrow Skin Patient | Eval Patient | Eval Spleen Testes Thymus Thyroid Urinary bladder wall Remaining organs Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Creating CT and PET Dose Reports Effective Date: 07/25/2018 | Quality HOOD05162002915465 Note: When either the PET Image Statistics or PET Dose Report is enabled during the use of Study Split for examinations with multiple accession numbers, the PET Image Statistics and/or PET Dose Report will not be assigned an accession number. To avoid this situation, ensure that the PET Image Statistics and PET Dose Report are disabled when utilizing the Study Split feature. Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Creating CT and PET Dose Reports Effective Date: 07/25/2018 | Quality HOOD05162002915465 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018
- Creating CT and PET Dose Reports