Replacing Flex Inserts R1 500/1000T Video

This video provides instructions on how to replace R1 flex inserts on the Dimension Vista® System.

To begin cleaning, the reagent flex inserts for R1, release the center interlock an lift the center lid. Locate the reagent arm to flex. Insert an. Loosen the thumb screw. This will be removed and cleaned later in this procedure. Now let's locate reagent arm one. Remove the reagent arm one insert by inserting a long slender tool in one of the insert holes, an lifted out of the instrument. Replace with the insert from the spare parts kit. Verify the insert is correctly seated with the open end on the right. Now that you have replaced the R1 Flex insert, close the lid to access the software.