Changing Cuvette Wash Platform Filter Video
This video provides instructions on how to change a cuvette wash platform filter on the Dimension Vista® System.
Release the center cover interlock and raise the lid. Remove the cuvet washer shield and set it aside. Next, remove the wash platform filter that needs to be replaced by holding the bottom of the filter. An lifting up. Lift the filter out by clasping the bottom of the filter and sliding it out an off the metal fitting that holds the filter in place. Place the new filter on to the top metal fitting. Push the filter into place by sliding it down onto the bottom metal fitting and ensure the filter is seated properly. Replace the Cuvet washer shield and closed the center instrument lid.
Change Cuvette Wash Platform Filter Filter
- maintenance